Test Procedures

Overview of the Testing and Measuring Procedures

Allow the Cuvette and Reagent Cap for each sample to reach room temperature before use. If the
MyCare Insite Analyser has been disconnected from the power supply, reconnect to reach operating
temperature before testing.
To analyse a patient sample:
Prepare Cuvette and Reagent Cap - allow at least 10 minutes to come to room temperature
-
Place the RFID card provided in the test box in the RFID card well on top of the Analyser
-
Enter the required patient data
-
Prepare a patient sample according to the instructions in the MyCare Insite Test package insert
-
Insert the cartridge into the Analyser and start the test measurment by closing the door
-
At the end of the measurement:
The door will open; remove cartridge and close door
-
The result will be displayed on the Touch Screen
-
The result will be saved in the Touch Screen's memory
-
If desired, export the result to an (optional) external computer, print the result with an (optional)
-
external printer (see page 21) or send it via email (see page 30)
Be sure to follow the detailed instructions for the test that are provided in the following sections and to
comply with the information provided in the package insert at MyCareInsite.com.
PI MCI-EUR Rev 05
26 / 50