Appendix
Portable and mobile RF communications equipment can affect medical electrical equipment.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
Medical devices may be affected by cellular telephones and other personal or household devices not intended for medical
facilities. It is recommended that all equipment used near the Vivatmo pro device comply with the medical electromagnetic
compatibility standard and to check before use that no interference is evident or possible. If interference is suspected or
possible, switching off the offending device is the normal solution, as is required in aircraft and medical facilities. Medical
electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided.

11.2 Compliant cables

WARNING
The use of accessories, transducers and cables other than those specified may result in increased
emissions or decreased immunity performance of the equipment or system.
The following list shows cables, transducers, and other applicable accessories with which Bosch Healthcare Solutions claims
EMC compliance.
Ethernet cable, 3 m, shielded
USB cable, 3 m shielded