WARNING
en
en
The Vivatmo pro is intended for use in professional healthcare facility environment. The customer or the user of Vivatmo pro
should assure that it is used in such an environment.
en
Medical devices may be affected by cellular telephones and other personal or household devices not intended for medical
facilities. It is recommended that all equipment used near the Vivatmo pro device comply with the medical electromagnetic
compatibility standard and to check before use that no interference is evident or possible. If interference is suspected or
en
possible, switching off the offending device is the normal solution, as is required in aircraft and medical facilities. Medical
electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to
en
the EMC information provided.
en

11.2 Compliant cables

en
WARNING
en
The following list shows cables, transducers, and other applicable accessories with which Bosch Healthcare Solutions claims
en
EMC compliance.
Ethernet cable, 3 m, shielded
en
USB cable, 3 m, shielded
54
Note that portable and mobile HF communication systems may interfere with this device.
Do not staple or use the device close to mobile phones or other devices generating electrical or
electromagnetic fields. This could result in malfunction of the medical device and may create a
potentially insecure situation. Portable RF communication devices (including peripherals such
as antenna cables and external antennas) are not to be used closer than 30 cm next to any part
of the Vivatmo pro system.
The use of accessories, transducers and cables other than those specified may result in increased
emissions or decreased immunity performance of the equipment or system.