Vivatmo me Content Your Vivatmo me ..........................................Welcome to Vivatmo me, the monitoring system for your respiratory disease Activate your Vivatmo me Intended use / Indications for use User interface Measurement of FeNO in exhaled breath ................................Attaching the disposable mouthpiece Performing the measurement...
Vivatmo me Activate your Vivatmo me 1. Open the battery compartment on the back of the Vivatmo me as operator. 2. Insert the 4 AAA primary batteries as shown on the inside of the battery compartment. 3. Remove the battery cover.
[ATS, 2011]. tic use) as an aid to monitor airway diseases whilst under Vivatmo me should only be used as directed in the Vivatmo me Instructions for use and as recommended by the care of their physician or healthcare expert. Vivatmo me measurement procedure requires patient’s coopera-...
Page 9
The following conditions can influence correct measurement results or cause malfunction or disturbance to the Vivatmo me and the disposable mouthpiece: • Smoking or tobacco consumption for at least 30 minutes before the measurement. • Eating or drinking shortly before the measurement.
Page 11
Your Vivatmo me Mouthpiece: Place your lips tightly around the mouthpiece and exhale into the Vivatmo me Display: Shows current and stored measuring data and device messages ON / OFF switch: Activate and deactivate the device ENTER button: Press this button to confirm the selection in the display ...
Measurement of FeNO in exhaled breath Attaching the disposable mouthpiece The Vivatmo me disposable mouthpiece chemically prepares the breath test sample. Human breath is cleaned from contaminants and stabilized to guarantee reliable test results. Use the disposable mouthpiece directly after opening the pouch.
For measuring the FeNO value, you have to exhale through the device with a constant breathing rate: 1. Turn ON the device, see “1.2 Activate your Vivatmo me”. 2. Attach a new disposable mouthpiece, see chapter “2.1 Attaching the disposable mouthpiece”.
Page 14
Measurement of FeNO in exhaled breath If the Vivatmo me device has been stored in warm and humid environment or not been in use for a longer time, the wrench symbol may appear and a reconditioning will start automatically to provide consistent measurement performance, see chapter “5.4 Reconditioning”.
Displaying stored measured values The Vivatmo me automatically stores up to 1,000 measured FeNO values identified by time, date, and ID number. The measured values are stored in descending order starting with the latest. You can retrieve the measured values anytime.
If you have further questions, contact the Vivatmo me service center. In order to access the advanced settings, press the ENTER button and the right ARROW button for 3 seconds.
1. Turn off the device. Wash hands thoroughly with soap and water. 2. Wipe the entire Vivatmo me with disinfectant wipes with maximum 30 % alcohol e.g. Mikrobac Tissues (Bode Chemie GmbH) or Sagrotan wipes (Reckitt Benckiser) or with a soft cloth, moistened with soap cleaning agents. Carefully wipe around openings.
• Always attach the protective cap when Vivatmo me is not in use. • Keep the Vivatmo me and its components in a clean, cool and dry place. The following conditions can cause damage and may cause malfunction to the Vivatmo me and the WARNING disposable mouthpiece: •...
Reconditioning If the Vivatmo me device has been stored in warm and humid environment or not been in use for a longer time, the wrench symbol may appear and a reconditioning is required to provide consistent measurement performance. In this case the wrench symbol appears during the regeneration process on the pacifier screen and the device automatically starts the reconditioning.
• Change the batteries if E-6 and the blinking battery symbol appears during reconditioning, see chapter “1.2 Activate your Vivatmo me”. • Change the disposable mouthpiece if the “Invalid mouthpiece” screen appears during recondi- tioning, see chapter “2.1 Attaching the disposable mouthpiece”.
Maintenance and Troubleshooting Troubleshooting When using the Vivatmo me, an error message may indicate a problem. The following table gives explanations of the error messages as well as recommendations for solving the problem. Display Description Solution Regeneration or reconditioning process Wait for the process to be completed.
Page 28
If you have any problems which you cannot solve with these Instructions for use or the device shows unexpected opera- tions, contact your distributor or the Vivatmo me customer service center at service@vivatmo.com. For the contact data see chapter “7 Manufacturer”. Please indicate the following information: •...
Technical information Device data Product description FeNO monitoring system Model Vivatmo me, catalogue number F09G100149 Measuring range 5 ppb to 300 ppb Accuracy ± 5 ppb below 50 ppb, ±10 % for ≥50 ppb, ± 15 % for ≥ 160 ppb...
Page 31
Further purchase: Vivatmo Oxycap mouthpieces (catalogue number F09G100152), Level 0 mouthpieces (catalogue number F09G100174). Limitations of the System: Exchange your Vivatmo me at the latest 3 years after manufacturing date. Information concerning candidates of the REACH regulation can be found at www.vivatmo.com.
Important information regarding electromagnetic compatibility (EMC) Hereby, Bosch Healthcare Solutions GmbH, declares that the radio equipment type Vivatmo me is in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: https://www.vivatmo.com/en/service-support/additional-information.html...
Technical information The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Bosch Healthcare Solutions GmbH is under license. Disposal of device, batteries and mouthpieces INFORMATION Delete the data on the device prior to disposal (see chapter 4.2).
WARNING Do not disassemble Vivatmo me or its components, and do not try to repair it by yourself. Any changes or modifications to Vivatmo me that are not approved by the manufacturer may cause malfunction and result in the warranty becoming invalid.