COILTYPE_D?
x
x
IEEE-488
USBTMC
Returns the default external current coil type used at power on and reset.
Parameters: None
Example: COILTYPE_D? returns C5500A50T
The default external current coil type is set to the 5500 50 turn coil.
COMM_ENABLE
x
x
IEEE-488
USBTMC
Enables or disables a remote interface port.
Parameters: 1. SERIAL, TELNET, GPIB, TMC
2. 1. (Boolean) ON or 1 for on, OFF or 0 for off.
Example:
COMM_ENABLE SERIAL, OFF
Disables the RS-232 Interface.
COMM_ENABLE TELNET, ON
Enables the telnet interface port.
COMM_ENABLE?
x
x
IEEE-488
USBTMC
Returns the active status of the interface port.
Parameters: 1. SERIAL, TELNET, GPIB, TMC
Response: ON if the interface is active, OFF if the interface is disabled.
Example:
COMM_ENABLE? SERIAL
Returns ON if the RS-232 interface is enabled/active, OFF if the RS-232 interface is disabled.
COMM_LOCK
x
x
IEEE-488
USBTMC
Locks out certain remote commands from executing on a specified remote port.
Parameters: 1. SERIAL, TELNET, GPIB, TMC
2. 1. (Boolean) ON or 1 for on, OFF or 0 for off
Example:
COMM_LOCK SERIAL, OFF Disables the lockout feature on the RS-232 Interface.
COMM_LOCK TELNET, ON
interface port.
x
x
x
RS-232
Ethernet
Sequential
x
x
x
RS-232
Ethernet
Sequential
x
x
x
RS-232
Ethernet
Sequential
x
x
x
RS-232
Ethernet
Sequential
Enables the command lockout feature on the Telnet
Overlapped
Coupled
Lockout
x
Overlapped
Coupled
Lockout
Overlapped
Coupled
Lockout
x
Overlapped
Coupled
Lockout
Calibrator
Remote Commands
59
Need help?
Do you have a question about the 5560A and is the answer not in the manual?