Optika Italy IM Series Instruction Manual page 12

Hide thumbs Also See for IM Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
4. Connect the cable from the external power supply to the
lamp housing. (Fig. 13)
5. Connect the power cable to the external power supply. (Fig.
14)
The input voltage for fluorescence power
supply is 110-240Vac.
Please use the standard power cable pro-
vided. Select suitable one when missing or
damaged.
Connect the power supply correctly, be sure
to have a good earth connection.
Before connecting the power cord, secure
the lamp housing cable to the power supply.
If the power cord is connected before, there
may be a risk of electrical shock.
Disconnect
installing or replacing the bulb.
The bulb has an anode and a cathode of
different sizes. Respect the polarity during
assembly.
6. In order to prevent possible damages from UV radiation,
mount the orange protection screen as shown. (Fig. 15)
all
electrical
cables
before
Page 12
F ig. 13
F
ig. 13
F ig. 14
F
ig. 14
F ig. 15
F
ig. 15