NIBP

NIBP Specifications
Characteristic
Method
Control
Auto Intervals
Turbocuf
Displayed Pressures
Systolic Range
Diastolic Range
Mean Range
Static Manometer Accuracy
Minimum Inflation Pressure
Maximum Allowable Pressure
Default Inflation Pressure
Normal Overpressure Limit (results in
up to 2 retries)
Single Fault Overpressure Limit
Leak Rate
Pulse Rate Range
Maximum Determination Time (with
retries)
Maximum Determination Time (no
retries)
Typical Determination Time without
Artifact
Welch Allyn
Oscillometric
Automatic and manual measurement control
1, 2, 3, 5, 10, 15, 30, and 60 minutes
Maximum measurements allowable in a 5-minute period
Systolic, Diastolic, and Mean plus on-screen manometer
Adult: 30 to 260 mmHg
Ped: 30 to 160 mmHg
Neonate: 25 to 120 mmHg
Adult: 20 to 235 mmHg
Ped: 15 to 130 mmHg
Neonate: 10 to 105 mmHg
Adult: 20 to 255 mmHg
Ped: 15 to 140 mmHg
Neonate: 10 to 110 mmHg
±3 mmHg
Adult: 100 mmHg
Ped: 80 mmHg
Neonate: 50 mmHg
Adult: 270 mmHg
Ped: 170 mmHg
Neonate: 132 mmHg
Adult: 160 mmHg
Ped: 120 mmHg
Neonate: 90 mmHg
Adult: 280 mmHg
Ped: 200 mmHg
Neonate: 141 mmHg
Adult: 308 mmHg
Ped: 220 mmHg
Neonate: 154 mmHg
After a 1 minute settling period, leak rate is ≤4mm/Hg over a 3-minute
period at 270 mm/Hg.
30 to 220 beats per minute
Adult: 4.5 minutes
Ped: 4 minutes
Neonate: 3 minutes
Adult: 3 minutes
Ped: 2 minutes
Neonate: 1.5 minutes
30 to 45 seconds
142
Specification
Propaq Encore Reference Guide
NIBP