Functional Verification

NIBP

1. Connect the cuff, digital pressure meter, bulb, and the Encore together with the cuff calibration
kit. Using T-connectors from the kit, connect the cuff, pressure meter, and bulb.
2. Connect this NIBP test setup to the pressure transducer at PT1 on the Analog board.
3. Wrap the adult cuff around the black tube volume or a cylindrical object about the size of an
adult's arm.
4. Press SETUP > MORE > MORE > SERVICE > YES > NIBP TEST > NIBP CAL.
Note
The NIBP CAL mode holds pressure for 10 minutes. If performing steps 4 through 9
takes longer than 10 minutes, press NIBP CAL again.
5. Close the valve on the bulb and inflate the cuff to 250 mmHg as shown on thepressure meter.
6. Check that the readings (PR1 and PR2) on the monitor's display are 250 ±3 mmHg (PR1) and
250 ±15 mmHg (PR2), as show in the figure on page 32.
19:00:25
NIBP TEST
CNTS
PR1:
578
PR2:
338
OSC:
1600
LEAK:
10
BLEED:
AUTO PMP
PUMP TEST
7. Manually reduce the pressure as indicated on the pressure meter to each of the levels shown in
the following table, and check the displayed value of PR1 to be within the tolerance listed in
the following table.
32
BATTERY: 8.1 VOLTS
mmHg
DAC
249.4
130
259
mmHg
DIAG:
mmHg/min
PUMP:
NIBP CAL
Pressure
(mmHg)
270
250
200
Section 2-Functional Verification
VALVE
0
0
secs
ZERO
Tolerance
(mmHg)
±
4
±
3
±
2
Propaq Encore Service Manual