Optopol SOCT Copernicus REVO User Manual
Hide thumbs Also See for SOCT Copernicus REVO:
Table of Contents

Advertisement

Quick Links

Manufacturer:
OPTOPOL Technology Sp. z o.o.
ul. Żabia 42
42-400 Zawiercie
POLAND
www.optopol.com
info@optopol.com.pl
SOCT Copernicus REVO
User Manual
Software Version 7
User Manual Rev. A
0197
Make sure you read this manual before using
the instrument. Keep this manual in a safe
place so that you can use it in the future.

Advertisement

Table of Contents
loading

Summary of Contents for Optopol SOCT Copernicus REVO

  • Page 1 Manufacturer: OPTOPOL Technology Sp. z o.o. ul. Żabia 42 42-400 Zawiercie POLAND www.optopol.com info@optopol.com.pl SOCT Copernicus REVO User Manual Software Version 7 User Manual Rev. A 0197 Make sure you read this manual before using the instrument. Keep this manual in a safe...
  • Page 2 Date of issue of manual: December 2016 Date of review: December 2016 For Customer Service or Technical Service contact your authorized dealer of OPTOPOL Technology Sp. z o.o. Contact details are available on www.optopol.com. SOCT Copernicus REVO User Manual Version 7.0 rev. A...
  • Page 3: Table Of Contents

    SOCT C REVO ..............22 OPERNICUS APPLICATION STRUCTURE PATIENT WINDOW ......................23 ......................... 24 ATIENT LIST VIEW ....................24 EGISTERING NEW PATIENTS ......................25 DITING PERSONAL DATA ....................... 26 NREGISTERING PATIENTS SOCT Copernicus REVO User Manual Version 7.0 rev. A...
  • Page 4 ....................... 52 XTERNAL FIXATION ..................52 HORIORETINAL ITREORETINAL MODE ........................ 53 XAMINATION TIPS RESULT REVIEW ......................... 55 ......................55 YPE OF 9.1.1 ...................... 55 INGLE TAB SCREEN 9.1.2 ....................55 TAB SCREEN SOCT Copernicus REVO User Manual Version 7.0 rev. A...
  • Page 5 FULLSCREEN WINDOW ....................96 12.1 SLO ..............96 UNDUS RECONSTRUCTION EYE PREVIEW OR P 12.2 ........................ 97 MAGING TOOLS 12.3 ................... 97 ELECTION AND LAYERS EDITION 12.4 ................97 EASUREMENT TOOLS AND ANNOTATIONS SOCT Copernicus REVO User Manual Version 7.0 rev. A...
  • Page 6 17.1 SOCT ......................120 NETWORK 17.2 SOCT ) ..121 HARING EXAMINATION DATABASE FOLDER INDOWS STORAGE LOCATIONS 17.3 SOCT / SOCT PC ............. 123 VIEWING STATION CONFIGURATION 17.4 FAQ ..........................124 SOCT Copernicus REVO User Manual Version 7.0 rev. A...
  • Page 7 ENVIRONMENTAL CONDITIONS ................... 125 WARRANTY AND SERVICE .................... 126 UTILIZATION ........................ 126 TROUBLESHOOTING ....................127 SOCT Copernicus REVO User Manual Version 7.0 rev. A...
  • Page 8: Description Of The Device

    The raw data are processed by a PC to obtain images of eye cross-sections. The SOCT Copernicus REVO contains features including: Retina Thickness, Optical Nerve Head, Retinal Nerve Fiber Layer, Angle assessment, Cornea measurement.
  • Page 9: Operating Skills

    CAUTION: This manual does not provide guidance on interpretation of clinical results. The clinician must ensure that he or she has received appropriate medical training in such interpretation. Optopol Technology Sp. z o.o. cannot be held responsible for misdiagnosis of results.
  • Page 10: Instruction Manual Availability

    1.2.7 Instruction Manual Availability The SOCT Copernicus REVO Instruction Manual in PDF format is available on the computer with installed SOCT application. To open the file, press START  Application  SOCT  User Manual. The manual is included on USB flash drive delivered with the standard package of the SOCT.
  • Page 11: Technical Data

    Class 1 Degree of protection against electric Type B applied parts (chin rest, forehead rest, and shock shock grip) Degree of Protection against IPX0 ingress of water Mode of operation Continuous operation 11 / 132 SOCT Copernicus REVO User manual...
  • Page 12: Minimum Computer System Requirements

    Windows 7 Professional or Ultimate SP1 or later, 64- bits Minimum 80 GB CD/DVD Recordable recommended Screen resolution: Minimum 1920 x 1080 Communication ports 1 available USB 3.0 port 2 available USB minimum 2.0 ports Mouse Mouse Wheel 12 / 132 SOCT Copernicus REVO User manual...
  • Page 13: Safety

    Electrical and Electronic Equipment waste. Do not throw the product away with normal household waste at the end of its life. Sign of conformity with essential requirements – The Medical Device Directive 93/42/EEC Class 1 Laser Product Warning: Dangerous voltage General warning sign 13 / 132 SOCT Copernicus REVO User manual...
  • Page 14: Product Label

    Caution: Federal law restricts this device to sale by or on the order of a physician or practitioner. 3.2 Product Label The sample label of the SOCT Copernicus REVO is presented below. The following table describes the marks and indications on the Spectral Optical Coherence Tomographer - SOCT Copernicus REVO.
  • Page 15: Safety Standards

    The SOCT Copernicus REVO system complies with all requirements of the directive 93/42/EEC (MDD - Medical Devices Directive). The SOCT Copernicus REVO is type B applied parts (chin rest support and forehead support) and class I protection against electric shock.
  • Page 16 There are no user serviceable parts inside SOCT Copernicus REVO device. Light source is included inside the device. Removing the covers can be done only by authorized personnel. The maximum power of light radiation available outside the covers is less than 1000 microwatts providing safe operation of the device.
  • Page 17: Notes On Use

    3.4.2 After use After using the SOCT device, turn off the power, attach the objective lens cap to protect the objective lens from dust and place the dust cover over the device. 17 / 132 SOCT Copernicus REVO User manual...
  • Page 18: Unpacking And Installation

    NOTE: Check for any damage to the package. Every transportation box is equipped with the shockwatches. If any shockwatch is broken (red indicator) please contact OPTOPOL Technology and lodge a complaint to the carrier. SOCT Copernicus REVO set consists of the following components: 1. SOCT device, 2. Fixation adapter (optional), 3.
  • Page 19: Connecting Cables

    -device will go to base position. Turn it off and pack to the box securing with packing foams. 4.2 Connecting cables To connect the SOCT Copernicus REVO to a PC, you need one USB 3.0 cable shipped with the device: USB 3.0 cable to connect SOCT Copernicus REVO (type B) and PC (type A): Power Supply Cable: All sockets and plugs are different so it is not possible to connect plugs improperly.
  • Page 20: Device Connection

    4.3 Device connection The SOCT Copernicus REVO is connected to the PC using USB 3.0 cable. NOTE: The SOCT Copernicus REVO must be directly plugged into the mains with protective earth. Connect the PC, printer and monitor into the multiple portable socket-outlet, which is directly connected to the Isolating Transformer.
  • Page 21: Factory Default Calibration And Configuration

    4.4 Factory default calibration and configuration The system is delivered fully installed and configured by local Optopol representative. No user installation or configuration is needed. On the computer connected to the SOCT screen saver, power options on the computer have to be set as shown in the table.
  • Page 22: Soct Copernicus Revo Software

    Figure 1 Login screen Enter the User Login and Password then press [Login] to go into the application or [Setup] to change settings. SOCT Copernicus REVO software is divided into three different modules which can be easily accessed from tabs: PATIENTS Patients and exams list appears login in the application.
  • Page 23: Patient Window

    Figure 2 Patient tab Result It enables to preview and analyze taken examinations. Software version Version of application. Device Status and SN Device status (Ready, Busy or not connected) and Serial Number. 23 / 132 SOCT Copernicus REVO User manual...
  • Page 24: Patient List View

    Other fields are optional and can be left empty. Figure 4 Patient edit screen NOTE: For patients with refractive error bigger than -/+ 5D it is recommended to fill in refraction during adding patients to the system. 24 / 132 SOCT Copernicus REVO User manual...
  • Page 25: Editing Personal Data

    [OK] button to confirm changes. System will check the data once again and display a warning message if entered data are incorrect . Then will be done the same checking as during the registration of a new patient. 25 / 132 SOCT Copernicus REVO User manual...
  • Page 26: Unregistering Patients

    Eye, Analysis program, Scan width, Scan dimension is a number of A-scans and B-scans are presented. Right Click to open Exam menu List of patient’s examinations. Thumbnail of Click to expand the exam list scanned image Figure 5 Examination list 26 / 132 SOCT Copernicus REVO User manual...
  • Page 27: Delete Exam/S

    6.6 Delete Exam/s Right-click on exam thumbnail and select [Delete]. 6.7 Export examination The SOCT Copernicus REVO offers the ability to export examinations data in raw.opt format. Exported examination in .opt can be imported to Copernicus application. Export data procedure.
  • Page 28: Filter

    Retina, Disc, Anterior, Central – depending on the analyzing program are available different views. Program Raster, Radial, 3D, single B-scan etc. Users Filter patient by operator who made examination. Use [Clear all] to reset the filter settings and disable filtering. 28 / 132 SOCT Copernicus REVO User manual...
  • Page 29: Examination Window

    [Setting] button. FollowUp Allows to continue previously conducted examinations. Device automatically moved to the same eye and selects the same scan parameters. Eye selector The measurement unit moves to the patient's selected eye. 29 / 132 SOCT Copernicus REVO User manual...
  • Page 30 Auto C (coherence) gate button to automatically align the OCT image. For patients with refraction error bigger than -/+ 5D it is recommended to fill refraction value during adding patients to the system. Focus slider Compensate the patient refraction. 30 / 132 SOCT Copernicus REVO User manual...
  • Page 31 Double click on the refraction value resets value to 0.00D. When cursor is over focus control panel, mouse scroll enables to compensate refraction. Lines on the vertical and horizontal window indicate proper position of tomograms. 31 / 132 SOCT Copernicus REVO User manual...
  • Page 32: Selection Of Scan Pattern Mode

    Useful for peripheral observation. Protocol: Protocol allows to perform predefined set of exams of different type one by one. This option allows to shorten time used for scanning program selection. 32 / 132 SOCT Copernicus REVO User manual...
  • Page 33: Selection Of Scanning Program

    *OCT Angiography analysis is an optional feature that may not be activated on all instruments. If you do not have this feature and want to purchase it, contact local Optopol’s distributor. 33 / 132...
  • Page 34: Selection Of Protocol

    Angle. Scan programs: [Retina 3D], [Disc 3D], and [Anterior Radial], [Anterior B-scan] [Screening] - This examination set captures tomograms of the macula, disc and central region of the retina. Scan programs: [Retina 3D] [Disc 3D] [Central 3D] 34 / 132 SOCT Copernicus REVO User manual...
  • Page 35: Follow Up

    . Press Left or Right eye button to move the device to the desired patient’s eye. Figure 10. Device movement controls. Touch screen control is an option. Contact Optopol representative for availability. 35 / 132 SOCT Copernicus REVO User manual...
  • Page 36 When checked system will start acquisition of data automatically after auto alignment of eye structure. When the cursor is over Eye preview window: scrolling the mouse wheel moves the head back and forth. 36 / 132 SOCT Copernicus REVO User manual...
  • Page 37: Eye Preview

    Click to center the pupil when working distance is aligned Figure 12.Device aligns to the place of clicking on the preview. Properly align the pupil to start searching oct signal. Figure 13. Properly aligned measurement head position 37 / 132 SOCT Copernicus REVO User manual...
  • Page 38: Pslo Live Fundus Preview

    NOTE: In case of changing position of internal fixation it may be required to compensate the scaning head postion to avoid cutting signal on the edges of the iris. 38 / 132 SOCT Copernicus REVO User manual...
  • Page 39: Fixation Target Change

    Scan height. It is available for Raster only, Spacing – Distance between consecutive tomograms. For Raster scan user can modify the distance, for all other programs the system displays the value automatically, Scan angle – change angle of scanning, 39 / 132 SOCT Copernicus REVO User manual...
  • Page 40 The system calculates the total number of A-scans and the examination time. After the changes have been made, user can press [OK] button to transfer new scan parameters to SOCT Copernicus REVO or [Save] to save modified parameters as a new default value. NOTE: Scan patterns in every scan mode have different settings.
  • Page 41: Live Oct Preview

    [Auto C-gate] compensates position of object on the OCT live window preview (length of coherence gate) and [A-Focus] (Refraction compensation) buttons and sliders to the left help you improve the scan image quality and centre it vertically. 41 / 132 SOCT Copernicus REVO User manual...
  • Page 42: Conducting Examination

    It is also possible to click on the eye preview(left mouse button or touch screen press) to move the head to desired location. 42 / 132 SOCT Copernicus REVO User manual...
  • Page 43: Acquisition Modes Description

    [align XYZ head position, Optimize OCT signal (C-gate, Focus), align tomogram position]. Full auto mode is disabled Figure 18. Semi auto examination. 2. OCT signal will be placed between horizontal lines. Figure 19. Semi automatic examination mode. 43 / 132 SOCT Copernicus REVO User manual...
  • Page 44: Figure 20. Proper Quality And Position Of The Tomogram

    NOTE: In case the system is not able to keep proper position of retina (e.g. patient is shaking) operator has to switch off tracking and make examination manually. 44 / 132 SOCT Copernicus REVO User manual...
  • Page 45: Manual Mode

     Move position of the internal fixation target. Ask patient to follow fixation target, compensate shadow. The OCT cross-section should be visible in OCT live preview window. Drag the retina to move tomogram preview to correct position. 45 / 132 SOCT Copernicus REVO User manual...
  • Page 46: Scanning Programs Description

    Mute or disabled. If required, use the large fixation target. See chapter 8.8 >>. 7. Verify scan program and change to RETINA if required. 8. Follow the procedure depending on Acquisition mode. Figure 24. Proper Alignment of retina tomogram. 46 / 132 SOCT Copernicus REVO User manual...
  • Page 47: Central Examination

    5) In Semi Auto or Manual when OCT signal is aligned properly, after a while the pSLO fundus preview image of the optic nerve appears. 6) The Optic nerve head has to be in the centre of scanned area. Figure 26. Proper Disk examination alighnment 47 / 132 SOCT Copernicus REVO User manual...
  • Page 48: Anterior Measurement

    Use centre reflex from cornea to locate the scan in the middle of scanned window. Use vertical dashed lines as reference. Figure 27. Proper Anterior measurement alignment. 48 / 132 SOCT Copernicus REVO User manual...
  • Page 49: Figure 28. Single Angle Measurement Proper Alignment

    NOTE: Vertical dense line in the center of Cornea is a natural reflection of laser light and has no negative influence on measurement result. It can be used to locate the tomogram in proper position. 49 / 132 SOCT Copernicus REVO User manual...
  • Page 50: Wide Anterior Programs

    8.3.5 Wide Anterior programs The Anterior Chamber Adapter for the SOCT Copernicus REVO is an easy-to-install hardware attachment to allow wide scanning of anterior segment structure. The lens adapter is attached and removed by a trained operator. Note: When using the anterior adapter, do not move the head too fast and monitor proximity to patient in order to avoid incidentally hitting the patient’s eye with the Anterior...
  • Page 51: Figure 32. Anterior Adapter Mounting. Rotate To Lock

     Angle to Angle scan – Ask patient to look at the green cross. Place the scan in the middle of the iris. Use pSLO view and dashed vertical line on live OCT window for reference. Both angles have to be visible on the live OCT window. 51 / 132 SOCT Copernicus REVO User manual...
  • Page 52: External Fixation

    With the external fixation method, the patient uses the second eye to fixate on an external target light. The SOCT Copernicus REVO may be equipped with an external fixation target arm. It is attached at the top of the forehead support. Its position is set manually. When you select external fixation target, instruct the patient to look at the blinking target light at the end of the external fixation arm.
  • Page 53: Examination Tips

    You can get a proper examination position by scrolling the mouse wheel or dragging to the desired position. Figure 37. Improper-flipped tomogram preview. Figure 38. Shadow on tomogram, grab and drag towards left side. 53 / 132 SOCT Copernicus REVO User manual...
  • Page 54: Figure 39. Shadow On Tomogram, Grab And Drag Towards Right Side

    Then click on the pupil or use up/dw and right/left movements to place the white cross in the center of equally aligned pupil. Figure 40. Proper live eye preview while manual acquire mode. 54 / 132 SOCT Copernicus REVO User manual...
  • Page 55: Result Review

    9.2 Types of Analysis 9.2.1 Single Retina analysis This shows the tomogram image of the macula and analysis results of retinal thickness from Retina 3D scan. The are available two analysis view. 55 / 132 SOCT Copernicus REVO User manual...
  • Page 56: Posterior Analysis

    From the menu available by the right mouse button you can select the image to overlay. The following images are available: Fundus reconstruction, pSLO, Fundus photo, Import fundus photo. To change transparency level, turn the mouse wheel on the fundus reconstruction image. 56 / 132 SOCT Copernicus REVO User manual...
  • Page 57 White area on the retina thickness chart identifies the area that is not covered by normative data. e.g. scan length higher than 7mm and fovea not in the centre of a scanned area. 57 / 132 SOCT Copernicus REVO User manual...
  • Page 58 For any particular measurement, note that 2 out of 20 normal eyes (10%) will fall above or below green. DISCLAIMER: OPTOPOL Technology Sp. z o.o. takes no responsibility for diagnostic interpretation of OCT images. It is the clinician’s responsibility to make diagnostic interpretations of OCT scans.
  • Page 59: Both Eyes Tab

    Sectors Map shows either average, maximum, minimum retinal thickness (in microns) or volume (in mm ) in each area. From Map Sector Dimension checklist choose circle diameters 0.6 / 2.22 / 3.45 mm or 1 / 3 / 6mm. 59 / 132 SOCT Copernicus REVO User manual...
  • Page 60 Blue dashed line shows axis used for asymmetry analysis. To modify the axis, place the cursor over the line, press and hold the left mouse button and move line to the desired position. 60 / 132 SOCT Copernicus REVO User manual...
  • Page 61 Sectors divide the elliptical annulus of the Thickness Map into 6 regions: 3 equally sized sectors in the superior region and 3 equally sized sectors in the inferior region. Dimensions of sectors: vertical inner diameter 1 mm and vertical outer diameter 4.2 mm, horizontal inner 61 / 132 SOCT Copernicus REVO User manual...
  • Page 62 NOTE: The user should visually evaluate the image to determine if the segmentation lines are correctly finding the analyzed boundaries. DISCLAIMER: OPTOPOL Technology Sp. z o.o. takes no responsibility for diagnostic interpretation of OCT images. It is the clinician’s responsibility to make diagnostic interpretations of OCT scans.
  • Page 63: Comparison

    The user can also change the dimension of the measurement rings. There are available two options: rings: 1; 3 and 6 mm (standardized ETDRS testing), rings: 0,6; 2,22 and 3,45 mm. 63 / 132 SOCT Copernicus REVO User manual...
  • Page 64: Progression

    Information table Trend plot for all zones in exams Figure 5 Follow up of the selected retina examination Consolidated plot shows the differences in each zone on one plot. 64 / 132 SOCT Copernicus REVO User manual...
  • Page 65 The rightmost column is the rate of change. The rate value appears only if the time interval between baseline and next exams is at least three months. Background colors are color-coded according to the normative database. 65 / 132 SOCT Copernicus REVO User manual...
  • Page 66 NOTE: Pay attention when you evaluate the tomograms on the Quick Printout if scanned widths were different between examinations. This shows the tomogram image from the macula up to the optic disc, and analysis results of retinal thickness. 66 / 132 SOCT Copernicus REVO User manual...
  • Page 67: Optic Nerve Head Analysis

    This view shows Fundus reconstruction with preview of selected tomogram, table with ONH Data information, RNFL thickness map in scan area and RNFL thickness graph. NFL thickness NFL Significant ONH table TSNIT profile thickness Tomogram preview TSNIT sectors Figure 41 Disc 3D [Single] tab 67 / 132 SOCT Copernicus REVO User manual...
  • Page 68 Any manual modification of recognition causes ONH Data calculation. DDLS section shows the DDLS stage, which is calculated from disc and rim dimensions. Calculated value is marked by yellow frame moving across the scale. 68 / 132 SOCT Copernicus REVO User manual...
  • Page 69 TSNIT SD are displayed above. DISCLAIMER: OPTOPOL Technology is not offering and does not offer advice, instruction in the diagnosis and interpretation of SOCT Copernicus REVO images. It is the clinician’s responsibility to make diagnosis and interpretation of OCT scans.
  • Page 70: Figure 42 Disc [Both] Tab

    Displayed map to overlay on the fundus reconstruction can be selected from the list box: NFL Significant NFL Deviation NFL thickness To change transparency level, turn the mouse wheel over the object. 70 / 132 SOCT Copernicus REVO User manual...
  • Page 71 TSNIT Average Standard Deviation Standard deviation of retinal thickness of TSNIT region Symmetry Average value of NFL thickness of TSNIT region (μm) Inter Eye Symmetry of both eyes relating to TSNIT region 71 / 132 SOCT Copernicus REVO User manual...
  • Page 72: Figure 43 Progression 3D Disc

    NFL Significance: Color map which shows result by comparing NFL thickness with the normative database NFL Deviation: Color map which shows differences between NFL thickness and the normative database. Disc contour blue color, Cup contour green color and measurement circles are shown. 72 / 132 SOCT Copernicus REVO User manual...
  • Page 73 The NFL parameter contains a difference column (from baseline) and rate of regression slope (the rate value appears only if the difference between baseline and next exams is at least three months). 73 / 132 SOCT Copernicus REVO User manual...
  • Page 74: Central Examination Analysis

    The fundus reconstruction image is created from all A-scans made in the scanned area. From the menu available by the right mouse button you can select the image to overlay. The following images are available: Fundus reconstruction, pSLO, Eye preview, 74 / 132 SOCT Copernicus REVO User manual...
  • Page 75: Scan Programs Results Review

    Anterior: Raster Single B-scan Radial Anterior Wide (with adapter): Raster Single B-scan Radial For more detailed description of above scan modes refer to chapter7.2 Selection of scanning program TU LINK DO 7.2 75 / 132 SOCT Copernicus REVO User manual...
  • Page 76: Single B-Scan Examination Review

    Single scan Layers display selection Figure 45. Single Scan view. Both Eyes Eye indicator and examination date Graph selection Figure 46. Single B-scan both eyes review. Comparison Figure 47. Single B-scan comparison window. 76 / 132 SOCT Copernicus REVO User manual...
  • Page 77: Raster Examination Results Review

    Raster examination result provides averaged image with enhanced resolution. Onew/Five tomograms are displayed depending on scanned region. Below examples of views: Single scan Figure 49. Retina raster, single scan view. Figure 50. Central raster single scan view. 77 / 132 SOCT Copernicus REVO User manual...
  • Page 78: Figure 51. Raster Both Eyes View

    Both eyes Figure 51. Raster both eyes view. Comparison Figure 52. Raster scan comparison. Progression Figure 53. Raster progression view. 78 / 132 SOCT Copernicus REVO User manual...
  • Page 79: Radial Examination Results Review

    Below example of single view: Figure 55. Cross scan horizontal B-scan view. Figure 56. Cross scan vertical B-scan view. Both eyes, Comparison, Progression displays are adequate to previously presented examples. 79 / 132 SOCT Copernicus REVO User manual...
  • Page 80: Visualization

    [Peeling]: Marked layers will be peeled during movement of red ball over tomograms. Operator can separately peel each layer. Easiest way to restore view is to use [Reset] button. [Select All]: Mark this item selects all layers. 80 / 132 SOCT Copernicus REVO User manual...
  • Page 81: Operation Panel

    [Surface] - It display non-transparent color-coded layers on the surface of each retina layer. [Fundus photo] – Option enabled only whenever a fundus image has been imported to the examination. Option shows common part of the imported image fundus image on the surface of retina. 81 / 132 SOCT Copernicus REVO User manual...
  • Page 82 Choose one direction to start animation slicing. It is possible to rotate reconstruction during slicing. To finish click [Stop] button which appears instead of play. [Save as] – Saves currently displayed picture in JPEG format. 82 / 132 SOCT Copernicus REVO User manual...
  • Page 83: Solid] View

    [Peeling]: Controls selected layers by using the peeling ball. The position of the layer is maintained when this is deselected. [Select All]: Clicking this item selects all layers. 7. Operation table Display operation table allows user modify the initial view. 83 / 132 SOCT Copernicus REVO User manual...
  • Page 84: Volume View]

    10.5.5 [Volume view] Shows the surface, layers of the retina and choroid as semi-transparent images. Figure 58 Volume view 84 / 132 SOCT Copernicus REVO User manual...
  • Page 85: Anterior Segment Analysis

    11 ANTERIOR SEGMENT ANALYSIS The SOCT Copernicus REVO was designed to posterior and anterior imaging. To obtain corneal and anterior segment images use Anterior Scan programs. Anterior scans for 3-5 mm width do not require anterior lens. Wide anterior scan with maximum 16 mm scan width requires to install anterior adapter on the objective lens.
  • Page 86: Figure 61 Sclera And Cornea Wide Anterior 11 Mm Scan

    Figure 61 Sclera and Cornea Wide Anterior 11 mm scan Figure 62 Angle to Angle view. Sample of wide anterior B-scan. 86 / 132 SOCT Copernicus REVO User manual...
  • Page 87: Anterior Radial

    It shows location of scan on the image. It can be selected from the list: Eye preview – captured image from front cameras pSLO – psudo SLO image 2. Information table Table displays summarizes the pachymetry data. 87 / 132 SOCT Copernicus REVO User manual...
  • Page 88 88 / 132 SOCT Copernicus REVO User manual...
  • Page 89 NOTE: Cornea analysis tab is available only when the system classifies the scanned object as Cornea scan. See letter in the top left corner. To change scan type go to Tomogram tab-> Menu -> Change Analysis type. Anterior Analysis type window will appear. Select [Cornea] type. 89 / 132 SOCT Copernicus REVO User manual...
  • Page 90: Both Eyes] Tab Screen

    This screen shows the analysis results comparing two examinations of one eyes on the same side in the same scan mode, from different dates. Exam info Eye preview Tomogram Thickness Measurement Difference map table 1. Eye preview 2. Thickness map 90 / 132 SOCT Copernicus REVO User manual...
  • Page 91: Progression] Tab Screen

    Tomograms allows evaluate morphology of scanned tissue 2. Pachymetry map The values overlay on the color coded thickness map. Values corresponding to the map are shown on the corneal grid. 3. Epithelium map. 91 / 132 SOCT Copernicus REVO User manual...
  • Page 92: Figure 65 Anterior Progression Tomogram View

    [S-I]: Difference between S sector and I sector within the corneal grid 11.1.4.2 Tomogram view Tomogram view allows to analyze morphology of the anterior structure. Type of view Thickness graph Eye preview Tomogram Latest exam Figure 65 Anterior Progression Tomogram view 92 / 132 SOCT Copernicus REVO User manual...
  • Page 93: Edit Anterior Surface

    Hold Left mouse button to draw boundaries Hold Right mouse button to delete boundaries Figure 66 Edit Anterior layer 93 / 132 SOCT Copernicus REVO User manual...
  • Page 94: Aod Measurement

    Angle Opening Distance of 750 micron. TISA: Trabecular Iris Segment Area. NOTE: Only when border air-anterior surface is correct are AOD and TISA measurements accurate. Verify recognition correction before judging the Anterior Angle morphology. 94 / 132 SOCT Copernicus REVO User manual...
  • Page 95: Angle Measurement Tool

    11.1.8 Tomogram review analysis This shows the tomogram image and the define thickness of the specified part. The compatible scan modes are [Raster], [B-scan], [Cross] and [Radial], . The tomogram images can be averaged. 95 / 132 SOCT Copernicus REVO User manual...
  • Page 96: Fullscreen Window

    Figure 67 Full screen tomogram 12.1 Fundus reconstruction, eye preview or pSLO Drop-down list box allows to toggle between fundus reconstruction, eye preview and pSLO. Figure 68 The choice of eye preview. 96 / 132 SOCT Copernicus REVO User manual...
  • Page 97: Imaging Tools

    Operator can also choose the color which the particular marking will be displayed on tomogram. Additional two buttons allows to delete single marking or all markings on particular tomogram. 97 / 132 SOCT Copernicus REVO User manual...
  • Page 98: Brightness And Contrast Adjustment

    12.6 Full screen mode exit Full screen mode exit – the button makes possible full screen mode exit and going to exam analysis. The same result is achieved by double clicking on tomogram. 98 / 132 SOCT Copernicus REVO User manual...
  • Page 99: Import Fundus Images

    Magnifier tool Fundus reconstruction Imported image Figure 74 Fundus registration screen Registration between the OCT fundus reconstruction image and imported retinal image may be review by changing transparency over result review window. 99 / 132 SOCT Copernicus REVO User manual...
  • Page 100: Examinations Correlation

    Use the mouse scroll as needed, to change the transparency to see more of Exam Image or Baseline Image. By moving the scroll back and forth, you can see if blood vessels or other features from one image align with the identical features in the other image. 100 / 132 SOCT Copernicus REVO User manual...
  • Page 101: Tomogram Window Manipulation

    Figure 76. Tomogram display options window. Double click on tomogram preview changes view to full screen and allows to edit layers recognition in the same way as in Retina examination viewing tab. 101 / 132 SOCT Copernicus REVO User manual...
  • Page 102: Edition Of Recognized Layers

    Correction can be used for fovea and disc analysis. It is enabled only after analysis is done. [Reanalyze] function is used to recover default recognition of layers. It is helpful in reanalysis of examinations is taken with previous software versions. 102 / 132 SOCT Copernicus REVO User manual...
  • Page 103: Angio Retina

    At lower left, an OCT en face image, next depth coded color map, at lower right tomogram B-scan image. Each angiogram has red (horizontal) lines that indicate the current B-scan location. You can drag these lines to display tomogram from desired location. 103 / 132 SOCT Copernicus REVO User manual...
  • Page 104: Angiogram Object

    IS/OS or RPE Therefore the angiogram images including IS/OS or RPE layer appear to be similar to angiogram of blood vessel structure of the inner retina. This effect is called Projection Artifact. In a healthy eye there are no blood vessels in the outer retina 104 / 132 SOCT Copernicus REVO User manual...
  • Page 105: Enface Window

    Additionally it is possible to modify position of boundary layers which creates angiogram view. Select boundary from the list box. Grab desire layer and move to desire position. It possible to change depth position of any boundaries. 105 / 132 SOCT Copernicus REVO User manual...
  • Page 106: Setup Window

    This tab allows the user to enter clinic details, change language or software look skin. Language selection Skin selection Enter name Enter address Select to load new logo Enter contact details Confirm [OK] or Cancer changes Figure 79.General setup window. 106 / 132 SOCT Copernicus REVO User manual...
  • Page 107: Database

    Don’t mark SOCT folder location only. Figure 82. Example of connection error. This message appears in case of wrong path indication or mistake in folder name Figure 83. Lack of directory error. 107 / 132 SOCT Copernicus REVO User manual...
  • Page 108: Storage

    (all folders should be shared and visible to all users) 15.3 Storage SOCT software allows to locate the database in various combined folders. It is possible to add more space for data storage if required and just indicate additional folder. 108 / 132 SOCT Copernicus REVO User manual...
  • Page 109 “Current” will be a storage for new examinations taken by the device. In case of viewing station connections storage locations should be shared in the network. For details see Network connection configuration 109 / 132 SOCT Copernicus REVO User manual...
  • Page 110: Users Accounts

    Confirm [OK] or Cancel changes Figure 88.Users tab. Enter user login Enter password User Name Retype the password User rights options Confirm [OK] or selection Cancel changes Figure 89. Users tab. Creating new user. 110 / 132 SOCT Copernicus REVO User manual...
  • Page 111: Preferences

    [Edit protocol] – allows to edit existing protocol [Delete protocol] – remove protocol from the list [Up and down arrow] - Move the position of selected protocol down or up on the list. 111 / 132 SOCT Copernicus REVO User manual...
  • Page 112 4. To delete the protocol, select desired Protocol from the Protocol list and click [Delete Protocol]. 5. To change the display order of Protocols or scan program, select desired item on the list, and then click [Up arrow] or [Down arrow]. 112 / 132 SOCT Copernicus REVO User manual...
  • Page 113: Parameters Tab

    When ‘Protocol’ is selected, the first protocol from the list is loaded. 15.5.3 Voice messages. In order to change voice settings press [Setup] button in Preferences tab. 'Voice message' tab contains options enabling customization of voice guide support or disabling it. 113 / 132 SOCT Copernicus REVO User manual...
  • Page 114 “Thank you, you can blink freely” – when system finishes acquiring examination. Shifting fixation for a disc Sound “Follow the green cross” – when system is changing fixation target during disc examination. 114 / 132 SOCT Copernicus REVO User manual...
  • Page 115: Results Settings

    Figure 94. Results review settings tabs. ‘Include IS/OS in thickness measurement’ – if this option is marked, PC adds thickness of Inner/Outer Segments of Photoreceptors to total value of thickness of retina. 115 / 132 SOCT Copernicus REVO User manual...
  • Page 116: Backup

    WARNING: Make sure there is enough free space on HDD/remote folder before performing backup process. In case this window appears please select backup folder location. Or check external HDD connection. Figure 96. Backup error example. 116 / 132 SOCT Copernicus REVO User manual...
  • Page 117: Maintenance And Cleaning Procedure

    WARNING: The main lens of the device should not contact the patient's eye or face. NOTICE: It is not allowed to make any modifications of the SOCT Copernicus REVO. During the whole period of use, SOCT Copernicus REVO does not need any special maintenance procedures.
  • Page 118: Software Maintenance Activities

    System administrator should check regularly for new updates at least once a month. 3. Installed antivirus software should be updated at least once a week. 4. For OPTOPOL's software upgrades and patches OPTOPOL will inform users through its distributors. 16.3...
  • Page 119: Ordering Consumables

    The fuses are located in power supply socket at the back of the device. In order to replace the fuses, unplug the power supply cable, press the small plastic lever and pull the fuses casing. 119 / 132 SOCT Copernicus REVO User manual...
  • Page 120: Network Connection Configuration

    Figure 97. Networking possibilities scheme. Networking functionality restrictions When software is configured as Client, there are some limitations applied to its functionality: Only client connected to SOCT device can perform new examinations. 120 / 132 SOCT Copernicus REVO User manual...
  • Page 121: Sharing Soct Examination Database Folder(S) On Windows 7(Storage Locations)

    NOTE: If for security reason this option cannot be activated ask network administrator to set up users and access right to SOCT folder. Accept and Save Changes. Figure 98. Sharing options setting. 121 / 132 SOCT Copernicus REVO User manual...
  • Page 122 Next configure share permissions. You should specify users which will have network access to SOCT folder. Grant them ‘Full control’. You can select ‘Everyone’ if you wish to grant access to everyone. Figure 101. Sharing options selection. 122 / 132 SOCT Copernicus REVO User manual...
  • Page 123: Soct Viewing Station Configuration/ Soct Pc

    17.3 SOCT viewing station configuration/ SOCT PC 1. In order to configure database client you need to install SOCT Copernicus REVO Software on each PC you want to use as a viewing station. You can find SOCT Copernicus REVO Software on the installation disc. To install it, run setup.exe.
  • Page 124: Faq

    Question: I have no connection to remote database Answer: Verify procedure of connection according to user manual again. Make sure network settings are proper. Verify if Host name /IP is properly entered in all software applications connected to server. 124 / 132 SOCT Copernicus REVO User manual...
  • Page 125: Environmental Conditions

    (including the connected devices). The SOCT Copernicus REVO with the optional power table requires an area at least (1,5 m by 2,4 m) for installation and patient comfort during use.
  • Page 126: Warranty And Service

    About repairs If a problem cannot be solved even after taking the measures indicated in “Chapter Troubleshooting” ), contact Optopol distributor for repairs. When requesting repair, please provide the following information: Full Name of the device and Serial Number: number on the rating label.
  • Page 127: Troubleshooting

    (in this case in right direction) on Vertical window right to obtain equal image saturation. The tomogram images have good quality, but the vertical tomogram image (right image) is diagonal. See A3 127 / 132 SOCT Copernicus REVO User manual...
  • Page 128 Protect physically USB plug from any loose. 3. Unplug and plug again the USB cable. Make sure there is no play (slack) on the port. Wait a while until system recognizes the devices again. 128 / 132 SOCT Copernicus REVO User manual...
  • Page 129 Q10: During Acquisition of tomograms OCT live windows become black. A10: See the status of the device in left bottom corner of the OCT image. In case of error restart application. If the problems occurs again see A9. 129 / 132 SOCT Copernicus REVO User manual...
  • Page 130 Figure 34. Angle to angle scan proper alignment..............52 Figure 35. Difference in Chorioretinal and Vitreoretinal scanning mode......52 Figure 36. Figure 30. Properly aligned retina between dashed lines........53 Figure 37. Improper-flipped tomogram preview..............53 130 / 132 SOCT Copernicus REVO User manual...
  • Page 131 Figure 7 Regulacja jasności i kontrastu ................... 98 Figure 67 Fundus registration screen ..................99 Figure 68 Exam registration screen ..................100 Figure 69. Tomogram display options window..............101 Figure 70 Single Angiography view ..................103 131 / 132 SOCT Copernicus REVO User manual...
  • Page 132 Figure 96. Advanced sharing settings change............122 Figure 97. Sharing options selection................122 Figure 98. Starting software and entering Setup............. 123 Figure 99.Remote database configuration..............123 Figure 100. Examination database storage location configuration........124 132 / 132 SOCT Copernicus REVO User manual...

Table of Contents