SOCT Copernicus +
Service Manual
FOR DISTRIBUTORS USE ONLY
Rev 2.7
(30-01-2014)
0197

Advertisement

Table of Contents
loading

Summary of Contents for Optopol SOCT Copernicus +

  • Page 1 Manufacturer: OPTOPOL Technology Sp z o.o. ul. Żabia 42 42-400 Zawiercie POLAND www.optopol.com info@optopol.com.pl SOCT Copernicus + Service Manual FOR DISTRIBUTORS USE ONLY Rev 2.7 (30-01-2014) 0197...
  • Page 2: Table Of Contents

    CONTENTS Introduction.......................... 5 Contact to Optopol Technology ..................... 5 Safety standards ........................6 Maintenance and cleaning ....................9 Device Overview ........................9 Technical Data ........................10 Unpacking and Installation ....................11 Unpacking SOCT Copernicus+ ....................11 Connecting cables ........................12 Device connection diagram ....................
  • Page 3 12. Electronics .......................... 80 12.1 Device block scheme ......................80 12.2 Tools needed to replace electronic components ..............81 12.3 Fuse exchange ........................81 12.4.Exchanging switching power supply..................84 12.4.1.Exchanging switching power supply................... 84 12.4.2.Exchanging transitional board 151-525................86 (For devices from S/N: 154012/I to S/N: 154196/L).
  • Page 4 13.3 Defocusing of spectrometer (too wide peak) ..............125 13.4 Too low spectrum on skantest ................... 128 13.5 Too high spectrum on skantest ..................130 13.6 Checking and adjusting power on objective lane ............... 131 13.7 Setting fixation point to the centre of the cross ..............135 13.8 Checking and adjusting the polarization ................
  • Page 5: Introduction

    For online support use program: TeamViewer (http://dane.optopol.com http://www.teamviewer.com/download/ In case when distributor decides send back the device to Optopol Technology Sp z o.o.. have to fill RMA form and send to service@optopol.com.pl in order to assign the RMA number. SOCT Copernicus PLUS - Service Manual...
  • Page 6: Safety Standards

    3. Safety standards SOCT Copernicus + system can be connected only to power supply socket equipped with properly connected grounding pin. The SOCT Copernicus+ system complies with all requirements of the directive 93/42/EEC (MDD - Medical Devices Directive) The SOCT Copernicus+ is Type B applied parts, class I protection against electric shock. Usually SOCT Copernicus+ is a part of larger Medical System, which consists of many other medical and non- medical devices.
  • Page 7 SOCT Copernicus PLUS - Service Manual...
  • Page 8 SOCT Copernicus+ has been classified as CLASS 1 LASER DEVICE Important! If the whole Medical System has been installed by Manufacturer or Authorized Personnel of SOCT Copernicus+, the Manufacturer guarantees correct installation and compliance with all required standards and directives, In case when installation was made not by Manufacturer or Authorized Personnel, Manufacturer of SOCT Copernicus+ takes NO responsibility for any problems or risks that could be created by incorrect connection and violation of safety standards,...
  • Page 9: Maintenance And Cleaning

    4. Maintenance and cleaning Warning! All maintenance activities can only be made when device is turned off and unplugged from power supply socket b) No user serviceable parts inside device. Any covers can be removed only by authorized service staff. During the whole period of use, SOCT Copernicus+ does not need any special maintenance procedures.
  • Page 10: Technical Data

    6. Technical Data Signal source Super Luminescent Diode (SLED), 830 nm (±2%) Optical power 1000 µW (950 - 1020 µW) Signal analysis Spectral domain Optical Coherence Tomography Scanning speed up to 27’000 measurements per second Scanning programs Single B-scan 3D visualization Asterisk Animation scan (selected number of B-scans in the same place at a retina)
  • Page 11: Unpacking And Installation

    7. Unpacking and Installation NOTICE!: Every transportation box is equipped with the shockwatches. If any shockwatch is broken (red indicator) please contact Optopol Technology Sp. z o.o. and lodge the complaint to the carrier. SOCT Copernicus+ set consists of following components: 1.
  • Page 12: Connecting Cables

    Now one can remove the cover by lifting it up gently. The SOCT’s Copernicus+ four legs are locked at two metal rails by clamps. There are only two removable clamps at a side. First remove screws from the clamps and then remove metal holders. This operation is shown in a figure below. Now shift SOCT Copernicus+ to the left and then lift it up.
  • Page 13 To connect the SOCT Copernicus+ to a computer, you need five cables shipped with the device: Camera link cable – connected to Copernicus + (not detachable), USB cables, connect to between Copernicus+ (type B) & PC (type A), Video/synchro cable connected with one plug Video (to PC) Synchro (BNC to PC) Video/Synchro to SOCT...
  • Page 14 First connect all signal cables described above and after that check the connections once again. In the next step connect power supply cables. NOTE: Regarding EMC (Electro-magnetic compatibility) standards all signal cables have to be put together. Picture on the next page shows how to connect all cables to the computer. SOCT Copernicus PLUS - Service Manual...
  • Page 15: Device Connection Diagram

    Device connection diagram The SOCT is connected to the PC using USB 2.0 and Video/Synchro cables. The SOCT must be directly plugged into the mains. Connect the PC, printer and monitor into the multiple portable socket-outlets, which is directly connected to the Separating Transformer. The transformer reduces power emission loss from the PC.
  • Page 16: Pc Installation

    8. PC installation Minimum system requirements (Starting on day 22.09.2010) In order to prepare computer for use with SOCT Copernicus+ you need: PC computer with operating system installed. Minimum system requirements: Processor: Intel® Xeon W3520 2.66 8MB/1066 QC CPU RAM: HP 4 GB (2x2GB) DDR3 –...
  • Page 17: Ni Frame Grabber Card And Winfast Card Installation

    NI Frame Grabber card and WinFast card installation Open the left side cover of a PC. Plug in the NI Frame Grabber into the PCI slot #2 then secure the card using a screw. Plug the WinFast Video card into the PCI slot #1 then secure the card using a screw. Close the PC cover.
  • Page 18: Drivers Installation

    Drivers installation 8.3.1 NI Vision Acquisition Software installation Turn the computer on. Insert a CD disc with SOCT Copernicus+ software to the CD-ROM. Open the CD and find the NI software files. Disc:\NI Vision Acquisition\VISIONACQ_CD1. Double click on the setup.exe. After initialization finishes, press Next>>.
  • Page 19 Select primary installation directory, it should be C:\Program files\National Instruments\ then press Next>>. Select features to be installed. Extend the icon NI-IMAQdx 3.1.2 and select “Do not install this feature”. Extend the icon NI-IMAQ I/O 2.1 and select “Do not install this feature”. SOCT Copernicus PLUS - Service Manual...
  • Page 20 Extend NI-IMAQ 4.0 and LabVIEW Support. Extend the icon LabVIEW 7.1 Support and select “Install this feature to a local drive”. Extend LabVIEW Real-Time 7.1 Support. Extend the icon LabVIEW Real-Time 7.1 Support and select “Install this feature SOCT Copernicus PLUS - Service Manual...
  • Page 21 to a local drive”. Press Next>>. Select “I accept the License Agreement(s)” then press Next>>. Press Next>>. SOCT Copernicus PLUS - Service Manual...
  • Page 22 The software is being installed now. If installation is interrupted and you are asked to indicate the Disc 2. Press the browse icon “..”. Find Disc 2 and press Select. SOCT Copernicus PLUS - Service Manual...
  • Page 23 Press Finish. Computer has to be restarted now, press Restart. After restart Hardware Wizard window will be opened then select No, not this time and press Next>> SOCT Copernicus PLUS - Service Manual...
  • Page 24 Select Install the software automatically (Recommended) then press Next>>. Press Finish, installation complete. Copy file with AVIIVA camera settings. Open SOCT installation CD and open Atmel_Aviva folder. SOCT Copernicus PLUS - Service Manual...
  • Page 25 Copy ATMEL AVIIVA SM2 CL 201x_rev2. File Locate the folder on HDD that contains the files “*.icd” using windows search.(menu Start>Search>files folders). Than right click on any of files and open containing folder. Paste setting file from CD to a found folder. If a pop-up window appears accept to replace existing file.
  • Page 26 Change the National Instruments driver settings. Double click at NI Measurement & Automation Software tools icon on Windows desktop. Extend My System, Devices and Interfaces, NI-IMAQ Devices, img0: IMAQ PCI-1428. Click right mouse button at channel 0. Extend Camera, Atmel and select ATMEL AVIIVA SM2 CL 201x_rev2.
  • Page 27: Usb Hub Drivers Installation

    8.3.2 USB HUB drivers installation Open the Device Manager. Start Control Panel System Properties Hardware Device Manager. Double click on USB Serial Port. Press Reinstall Driver… Select No, not this time then press Next>>. SOCT Copernicus PLUS - Service Manual...
  • Page 28 Select Install the software automatically (Recommended) then press Next>>. Press Finish. SOCT Copernicus PLUS - Service Manual...
  • Page 29: Winfast Tv Card Drivers Installation

    8.3.3 WinFast TV Card drivers installation Open a Device Manager. Start Control Panel System Properties Hardware Device Manager. Double click on Multimedia Controller. Press Reinstall Driver… SOCT Copernicus PLUS - Service Manual...
  • Page 30 Select No, not this time then press Next>>. Select Install the software automatically (Recommended) then press Next>>. Press Finish. SOCT Copernicus PLUS - Service Manual...
  • Page 31 Follow the same procedure with Multimedia Video Controller. SOCT Copernicus PLUS - Service Manual...
  • Page 32: Microsoft Visual C++ Redistributable Installation

    8.3.4 Microsoft Visual C++ Redistributable installation Insert a CD with SOCT Copernicus + software to the CD-ROM. Open the disc and double click on the vcredist_x86_vc2008.exe. Press Next>>. Select I have read and accept the license terms then press Install >>. SOCT Copernicus PLUS - Service Manual...
  • Page 33 Wait till then press Finish. SOCT Copernicus PLUS - Service Manual...
  • Page 34: Soct Software Installation

    SOCT Software installation Insert a CD with SOCT Copernicus + software to the CD-ROM. Open the CD and double click on the setup.exe. Choose language and press Press Next>>. SOCT Copernicus PLUS - Service Manual...
  • Page 35 Select destination location (Recommended location is D/SOCT). If you would like to select a different folder click Browse. Then press Next>>. Select start menu folder then press Next>>. Select Create a desktop icon then press Next>>. Press Install. SOCT Copernicus PLUS - Service Manual...
  • Page 36 The software is installing now. Press Press Yes. SOCT Copernicus PLUS - Service Manual...
  • Page 37 Installation complete press Finish. Connect the PC to SOCT run the software, perform test examinations, make sure the system is working properly. SOCT Copernicus PLUS - Service Manual...
  • Page 38: Troubleshooting

    9. Troubleshooting Mechanical troubleshooting Plug the Power cord and turn on the device See: Fuse Exchange, Does unit start? See: Electronic troubleshooting Protect handle is still In device, Have Coppernicus PLUS (see: Removing protect handle) positioned itself in the middle position? Can chinrest perform up and See:...
  • Page 39: Electronic Troubleshooting

    Electronic troubleshooting 1. Device does not 1.1. Protect handler is still in device. CCW undo protect handler with is on the turn on. back of device and remove it 1.2. Device is not plugged into Power Check the Power supply cable connection – supply grid.
  • Page 40 2.4. No feeding on board 151-506 Check board 151-506 (see: Board 151-506) If Any of the supplying voltage on board 151-506is invalid, check switch source power supply (see: Switch source power supply) 2.4. Main controller broken (151-506) Check 151-506 board, (see: Board 151-506) If board feeding is correct, and there is no...
  • Page 41 4.2. Fault of controlling of SLED. Turn on the SLED diode In Skantest program and check voltage between pins X10.2 X10.1 Voltage after Turing on the diode should be < 0.15V.next step is to turn off diode from skantest program and measure voltage again.
  • Page 42 7. No movement in 7.1 No Power supply or steering on Check feeding of board 151-517 check X,Y,Z axis board 151-517 connection of plugs X5,X6,X7 (see: Board 151-517) If engine board is properly feed, check Communications between boards 151-517 (plug X1 ) and 151-506(plug X2) (see: Board 151-506).
  • Page 43: Optical Troubleshooting

    (Setting fixation point to the centre of the cross Still wrong Problem solved  Check the polarization adjustments Still wrong Problem solved Reference lane filter adjustment  Still wrong Problem Back to the Optopol solved  SOCT Copernicus PLUS - Service Manual...
  • Page 44 BAD IMAGE QUALITY (Doubled, tripled, lots of noise) Check the spectrometer's adjustment Still wrong Problem solved  Swap setup.ini file Check the polarization Problem solved  adjustments Problem Back to the Optopol solved  SOCT Copernicus PLUS - Service Manual...
  • Page 45 TOO HIGH SPECTRUM IN SKANTEST PROGRAM Check and adjust power on the objective lane Problem Still wrong solved Adjust power Reference lane filter adjustments Problem solved SOCT Copernicus PLUS - Service Manual...
  • Page 46 TOO LOW SPECTRUM IN SCANTEST PROGRAM Check the spectrometer adjustments Still wrong Problem Adjusting power on objective lane solved  Still wrong Check the peak shape with Problem solved  crossed spectrometer Good Bad shape shape Reference lane filter Adjusting spectrometer adjustments SOCT Copernicus PLUS - Service Manual...
  • Page 47: Opening And Closing The Device

    2. BE CARREFUL WHEN YOU ARE OPENING THE SOCT PLUS. DO NOT REPLACE POSITION OF OPTIC COMPONENTS. If you move a position of optics probably you will have to send back the unit to Optopol Technology Sp. z o.o. for calibration. Especially do not touch the optic fibre (Red arrows).
  • Page 48 3. Run a program scantest_lt.exe, set chinrest in the lowest position. Set position of head: maximum to top, maximum to back and in a middle between left and right. Turn off the SOCT Copernicus. 4. Make sure, that blocking device is removed (look picture below). 5.
  • Page 49 6. Dethatch lens cover. Wipe your hands, so there is no slip, catch frame as showed below and pull towards to yourself. SOCT Copernicus PLUS - Service Manual...
  • Page 50 7. Undo screws holding cover slides. 8. Detach lens cover . SOCT Copernicus PLUS - Service Manual...
  • Page 51 Undo 4 screws holding upper cover. 10. After unscrewing all 4 screws take off the cover (see photos below): SOCT Copernicus PLUS - Service Manual...
  • Page 52 11. Unit without upper cover: 12. Disassembly of the upper cover, carrying frame. Firstly mark frame’s position according to its holder: SOCT Copernicus PLUS - Service Manual...
  • Page 53 13. Place soft paper towel under the frame in order not to scratch cover Soft towel 14. Undo 4 screws holding brackets of the frame (two on each side). 15. After that you can take off the carrying frame: SOCT Copernicus PLUS - Service Manual...
  • Page 54 16. Loosen 2 screws holding 2 frame supporters. Front view Rear view 17. Take out both supporters: SOCT Copernicus PLUS - Service Manual...
  • Page 55 19. Take out the chinrest and the forehead rest. This step can be done at the very beginning: 20. Undo the external fixation plug and undo screw holding the ground cable of forehead rest: SOCT Copernicus PLUS - Service Manual...
  • Page 56 21. To take off cover you have to undo 8 screws from underneath the unit. SOCT Copernicus PLUS - Service Manual...
  • Page 57 22. Place unit that its legs stand close to the edge of the table and undo all 8 screws. You need to feel the screw. It is placed on height of the connection of the covers. NOTICE: Device must stand on all 4 feet, to avoid damage. SOCT Copernicus PLUS - Service Manual...
  • Page 58 23. After unscrewing all 8 screws You can take off the cover: Set unit to the maximum back, and up position, and centre it in X axis. Also move chinrest to the maximum down position, b) Carefully lift the cover, c) While approaching the scanners turn cover around in order not to damage scanners.
  • Page 59: Closing Cover

    2. BE CARREFUL WHEN YOU ARE CLOSE THE SOCT PLUS. DO NOT REPLACE POSITION OF OPTIC COMPONENTS. If you move a position of optics probably you will have to send back the unit to Optopol for calibration. Especially do not touch the optic fibre! SOCT Copernicus PLUS - Service Manual...
  • Page 60 3. Mount cover on device. a) Place the cover on the top of the unit b, c) Manoeuvre cover not to touch optical fibres and polarisation, d) Turn cover in order not to damage scanners and it’s plugs, e) Mount cover on the unit. SOCT Copernicus PLUS - Service Manual...
  • Page 61 4. Screw in all 8 screws holding the cover. To attach cover you have to tighten 8 screws from underneath the unit. 5. Place unit that its legs stand close to the edge of the table and tighten all 8 screws.: 6.
  • Page 62 7. Tighten the screws holding frame supporters (remember to position it with markers made while disassembling the cover: Rear view Front view SOCT Copernicus PLUS - Service Manual...
  • Page 63 8. Mount the carrying frame: 9. Tighten 4 screws of carrying frame: SOCT Copernicus PLUS - Service Manual...
  • Page 64 10. Screw in the forehead rest grounding cable and attach external fixation plug. 1. Spring washer 2. Connector 3.Washer 4. Screw SOCT Copernicus PLUS - Service Manual...
  • Page 65 11. Before assembling forehead rest remember to place external fixation plug inside headrest. 12. Assembly the upper cover. SOCT Copernicus PLUS - Service Manual...
  • Page 66 13. Tighten 4 screws. 14. Attach cover lens and tighten the screws. 15. Attach lens cover. SOCT Copernicus PLUS - Service Manual...
  • Page 67: Movement Modules

    11. Movement modules 11.1.1 Chinrest module 1. Open the SOCT cover (See: Opening cover), 2. Mark chinrest position with marker. 3. Disconnect power supply cables from connector and engine. 4. Unscrew chinrest from base plate. SOCT Copernicus PLUS - Service Manual...
  • Page 68 5. Take out the chinrest. 6. Mark engine position using marker. 7. Unscrew engine from the base plate and lift it until plastic wheel reaches slide. SOCT Copernicus PLUS - Service Manual...
  • Page 69 8. Use tape to demobilize unit (according to protect micro switches from being destroyed). 9. Take out engine module with screw. 10. Assembly new rubber band to the engine unit: SOCT Copernicus PLUS - Service Manual...
  • Page 70: Chinrest Module Replace In Devices With Serial Number Higher Then 154229/L

    11. Reassembly chinrest following steps 3-8 in opposite order. 12. Close the cover (See: Closing the cover). 11.1.2 Chinrest module replace in devices with serial number higher then 154229/L. 1. Open the SOCT cover (See: Opening cover), 2. Mark chinrest position with marker. 3.
  • Page 71 4. Unscrew screw holding grounding cable. 5. Unscrew chinrest from base plate. 6. Take out the chinrest. 7. Reassembly chinrest following steps 3-8 in opposite order. 8. Close the cover (See: Closing the cover). SOCT Copernicus PLUS - Service Manual...
  • Page 72: Left-Right Movement Engine

    11.2 Left-right movement engine Tools needed to replace engine: - hexagonal tools size: 2, 2.5 - magnetic holder, 1. Read entire manual first before you start opening the SOCT PLUS. 2. Undo the engine screws and unscrew 2 screws holding engine frame to its supporter. Unscrew 4 screws holding engine to its frame.
  • Page 73: Front-Backward Movement Engine

    11.3 Front-backward movement engine Tools needed to replace engine: - hexagonal tools size: 2, 2.5 - magnetic holder -tweezers -side cutters -side cutters 1. Turn off the device. 2. Open the cover (see: Opening cover) 3. Unplug engine’s plug and cut plastic tightening band. Unplug Unscrew 4.
  • Page 74 5. Unscrew 4 screws holding engine to its frame. 6. To assembly new engine follow upper steps in opposite order. 7. Make sure that new engine’s rack is parallel to the toothed bar, if not – perform regulations! SOCT Copernicus PLUS - Service Manual...
  • Page 75: Up-Down Movement Engine

    11.4 Up-down movement engine Tools needed to replace engine: - hexagonal tools size: 2, 2.5 - magnetic holder -tweezers 1. Turn off the device. 2. Open cover (see: Opening cover). 3. Make sure that during engine exchange, you won’t change position of. 4.
  • Page 76 7. Remove bearing by unscrewing 2 screws and picking it out with tweezers. SOCT Copernicus PLUS - Service Manual...
  • Page 77 8. Unscrew 2 screws holding engine frame to its supporter. 9. Carefully take out the engine. 10. Unscrew 4 screws holding engine to its frame and assembly new engine. SOCT Copernicus PLUS - Service Manual...
  • Page 78 11. Place new engine on place of old one and mount it to the frame. 12. Make sure that new engine’s rack is parallel to the toothed bar, if not – perform regulations! 13. Reassembly bearing to the frame and place bearing guard back on place. SOCT Copernicus PLUS - Service Manual...
  • Page 79 14. Reassembly electronic board back to the frame by with 4 screws. 15. Check if polarisation haven’t changed (see: Checking and adjusting the polarization). SOCT Copernicus PLUS - Service Manual...
  • Page 80: Electronics

    12. Electronics 12.1 Device block scheme SOCT Copernicus PLUS - Service Manual...
  • Page 81: Tools Needed To Replace Electronic Components

    12.2 Tools needed to replace electronic components Pic. 2 Tools required for replacing electronics 1. Hexagonal key size 2. 2. Flat-head screwdriver. 3. Philips-head screwdriver. 4. Multimeter. 12.3 Fuse exchange Turn off the device– Lamp on switch don’t shine. Unplug the power supply cord. SOCT Copernicus PLUS - Service Manual...
  • Page 82 Unlock fuse cover from both sides. If necessarily use a thin flat head screwdriver. Remove fuse’s socket. SOCT Copernicus PLUS - Service Manual...
  • Page 83 Replace fuses. Plug fuse’s socket back on place. SOCT Copernicus PLUS - Service Manual...
  • Page 84: Exchanging Switching Power Supply

    12.4.Exchanging switching power supply. 12.4.1.Exchanging switching power supply. Unplug power cord from device. Open the cover (see: Opening cover). Undo two screws holding schurter filter. Move away filter from the switching power supply. Disconnect power plug from transitional board and disconnect AC voltage IN plug. Cut all tie-wraps marked on picture below.
  • Page 85 Undo screw holding PE cable. Undo four screws holding switching power supply. Undo four screws holding transitional board. 10. Take switching power supply out of device. 11. Fix new switching power supply with thermal grease. 12. Do screws holding transitional board. 13.
  • Page 86: Exchanging Transitional Board 151-525

    12.4.2.Exchanging transitional board 151-525. (For devices from S/N: 154012/I S/N: 154196/L). Unplug power cord from SOCT device. Open the cover (see: Opening cover). Undo four screws holding transitional board. Disconnect all plugs from transitional board. Take the transitional board out of the device. Fix new transitional board.
  • Page 87 Connect all plugs to the proper sockets according to below description. Transitional board 151-525 Socket Plug Description Not Connected +12V, CPU master 151-506 +12V, Camera AviivA Power supply +12V, Eye preview camera power supply Not Connected Not Connected +/- 24V, Galvo scanners power supply +/- 24V, Motors module 151-517 power supply Not Connected Not Connected...
  • Page 88: Exchanging Transitional Board 151-527 (For Devices From S/N: 154208/L)

    12.4.3.Exchanging transitional board 151-527 (For devices from S/N: 154208/L). Unplug power cord from SOCT device. Open the cover (see: Opening cover). Undo four screws holding transitional board. Disconnect all plugs from transitional board. Take the transitional board out of the device. Fix new transitional board.
  • Page 89 Connect all plugs to the proper sockets according to below description. Transitional board 151-527. Socket Plug Description Not Connected +12V, CPU master 151-506 +12V, Camera AviivA Power supply +12V, Eye preview camera power supply +/- 24V, Galvo scanners power supply X +/- 24V, Galvo scanners power supply Y +/- 24V, Motors module 151-517 power supply Not Connected...
  • Page 90: Exchanging Transitional Board From 151-526 (151-525) To 151-527

    12.4.4.Exchanging transitional board from 151-526 (151-525) to 151-527. (For devices from S/N: 154012/I S/N: 151207/L). Unplug power cord from SOCT device. Open the cover (see: Opening cover). Undo four screws holding transitional board. Disconnect all plugs from transitional board. Take the transitional board out of the device. Fix new transitional board.
  • Page 91 Cut old scanners power cable X7 and power cables from scanners side. Cut power cables from scanners side. SOCT Copernicus PLUS - Service Manual...
  • Page 92 10. Install new scanners power cable (white black blue red wire). Install the scanners power cables directly on main wire. SOCT Copernicus PLUS - Service Manual...
  • Page 93 11. Connect scanners power cables to scanners control boards. 12. Use special adapter to connect plug X16. SOCT Copernicus PLUS - Service Manual...
  • Page 94 X scanner voltage verification 14. Close the cover (see: Closing cover). It is necessary to contact with Optopol Technology Sp. z o.o. Service Division if the device have S/N: 154011/I; 154013/I; 154014/I; 154015/I; 154016/J; 154017/J; 154018/J. SOCT Copernicus PLUS - Service Manual...
  • Page 95: Main Controller Board 151-506

    12.5 Main controller Board 151-506 Board 151-506 is main controlling part of electronic systems in device, It controls all electronic systems in device and communicates with computer. X7 X6 SOCT Copernicus PLUS - Service Manual...
  • Page 96: Checking Correctness Of Feeding On Board 151-506

    12.5.1 Checking correctness of feeding on board 151-506 LED2 LED1 After turning on the power supply diode signalizing proper functioning of random parts of board should shine. If diodes LED 1 or LED 2 doesn’t shine check voltage on connectors X1 and X5. Tab.
  • Page 97: Exchanging Board 151-506

    12.5.2 Exchanging board 151-506 Unplug device from power socket . Open cover (see: Opening cover). Disconnect all plugs from board 151-506. Unscrew 4 mounting screws. Assembly new board. Plug all plugs according to scheme. SOCT Copernicus PLUS - Service Manual...
  • Page 98: Board 151-512

    12.6 Board 151-512 Destination of board 151-512 is controlling fixation diodes, servomechanisms of refraction. It is slave type board, it directly executes orders from main driver - board 151-506. SOCT Copernicus PLUS - Service Manual...
  • Page 99: Control Of Feeding On Board 151-512

    12.6.1 Control of feeding on board 151-512 After Turing on device, diode should light on board 151-512 informing of proper board feeding. LED1 Id diode LED1 stays off, check voltage on plug X1 Diode Plug Pins Voltage DC X1.1 – X1.2 11.7V –...
  • Page 100: Exchanging Of Board 151-512

    12.6.2 Exchanging of board 151-512 Plug device out of powering cord. Open cover (see: Opening cover). Unplug All plugs prom board 151-512. Unscrew 4 screws holding board. Detatch old board and attatch new one. Connect all plugs into proper sockets. SOCT Copernicus PLUS - Service Manual...
  • Page 101: Movement Control Board 151 - 517

    12.7 Movement control board 151 – 517 Board 151-517 is complete engine movement control board in axis X,Y,Z It’s slave type board controlled by the main controller board 151-506 Pic 4 Placement of connectors on board 151-517 SOCT Copernicus PLUS - Service Manual...
  • Page 102 Voltages on power supplying plug X1 Plug Pins Voltage DC X1.1 - X1.2 - (22V – 25V) X1.3 - X1.2 + (22V – 25V) Description of connectors on board 151-517 Plug Function Powering socket USB controller Y axis movement transoptor Z axis movement transoptor X axis movement transoptor Communication...
  • Page 103: Exchange Of Board 151 - 517

    12.7.1 Exchange of board 151 – 517 Detach power supplying cord. Open cover (see: Opening cover). Detach board 151-506 (see: detaching board 151-506). Detach all connectors from board 151-517. Unscrew 2 mounting screws. Attach new board. Connect all plugs to proper sockets. Continue exchange in reverse order to dethatching.
  • Page 104: Board 151-126

    12.8 Board 151-126 Board 151-123 is RS232-USB converter. It mediates during data exchange between board 151-506 and PC computer. 12.8.1 Checking correctness functioning of board 151-126 LED1 LED2 Checking communication – In program skantest try connecting to the device ,if connection is unavailable to be set, check if diode LED1 shines while attempting the connection.
  • Page 105: Exchange Of Board 151 - 126

    12.8.2 Exchange of board 151 – 126 Detach power supply cord from device. Open cover (see: Opening cover). Detach connector from board 151-126. Unscrew mounting screws. Detach old and attach new board. Connect plugs to proper connectors . SOCT Copernicus PLUS - Service Manual...
  • Page 106: Scanner Controlling Boards

    12.9 Scanner controlling boards Scanner powering control Diodes signalizing scanner’s status Both scanners controls boards are equipped with diodes. If diode shining colour is: Green- Scanners are turned on, and Works properly Red – Scanner error, system malfunction Orange – Scanners are turned off SOCT Copernicus PLUS - Service Manual...
  • Page 107: Exchange Of Scanner Driver Boards

    12.9.1 Exchange of scanner driver boards Unplug device from power supply. Open cover (see: Opening cover). Dethatch all plugs from boards. Unscrew 2 screws on each board. Detach old and attach new boards. Connect plugs to proper connectors. SOCT Copernicus PLUS - Service Manual...
  • Page 108: Exchange Of Spectrometer Servomechanism Controller Board 151-539

    12.10 Exchange of spectrometer servomechanism controller board 151-539 Unplug camera power supply plug. Unscrew with flat-headed screwdriver two screws attaching camera signal cable. SOCT Copernicus PLUS - Service Manual...
  • Page 109 Detach plug from controller. Unscrew 4 screws holding electronic board. If extreme detectors plates interrupts taking board out – turn CCW servomechanism engine to enlarge gap between plates and upper screws. SOCT Copernicus PLUS - Service Manual...
  • Page 110: Checking Sled Diode

    12.11 Checking SLED diode Proper connection of SLED diode. Proper SLED driver functioning is signalized by LED diodes shining green. SOCT Copernicus PLUS - Service Manual...
  • Page 111: Checking Aviva Camera Power Supply

    If power supply board works and camera connection is proper and diode stays off, most probably camera is broken – contact authorised service (see: Contact to Optopol Technology Sp. z o.o.). SOCT Copernicus PLUS - Service Manual...
  • Page 112: Exchanging Cl Cables

    12.13 Exchanging CL Cables In SOCT Copernicus PLUS unit, highly elastic CL cable should be used. There are two CL cables in Copernicus PLUS. Each cable is signed described- camera on unit’s side and Grabber on computer side. Due o install cable clamps (cable clamp N-4), it’s required to set mounting point on the cable. On cable measure 26 cm ±...
  • Page 113 Prepared cables mount to the base plate using M16x4 screws Note: Cable distance to the clamp (26 cm) should go through lower hole in the rear panel and should be connected to the CL camera slot. Propertly connected CL cable should look like on picture below: Cable can’t touch housing.
  • Page 114 Turn on the Skantest program and initialize sled + movement. Warm unit for 40 minutes When system is warmed perform spectrometer stability test: In scantest program use option [FAST] to adjust spectrometer. 1. Observe the band height 1 hour from adjusting spectrometer. Signal cannot drop more than 10%.
  • Page 115: Rear Panel

    12.14 Rear panel 1. Properly connected wiring of rear panel. All plugs pushed into sockets. Device turned on. SOCT Copernicus PLUS - Service Manual...
  • Page 116: Programming The Cpu

    12.15 Programming the CPU Before you start the programming procedure contact Optopol technology service department. Prepare the following information: serial number of the device taken from the scantest, or SOCT software patient tab(all numbers). Request programming files, it will be prepared and sent to you via e-mail.
  • Page 117 5. Turn the device on. 6. Run program aduc842prog.exe. Set speed of transmission of COM port on 9600. Choose proper COM port. SOCT Copernicus PLUS - Service Manual...
  • Page 118 7. Open COM port. Press „Otwórz”. Software will detect microcontroller ADUC842. 8. Delete the microcontroller’s firmware. Press „Kasuj program i dane”. Repeat that step 4 times. SOCT Copernicus PLUS - Service Manual...
  • Page 119 Load firmware. Press „Ładuj program” and choose prepared by Optopol .hex file. Software has to display message which inform about success of programming. “Programowanie zakończone sukcesem” . Then start the processor, accept proper work and close the program ”Koniec pracy”.
  • Page 120 10. Turn OFF SOCT. 11. Remove programming plug or resistor. 12. Connect synchro-video cabel. SOCT Copernicus PLUS - Service Manual...
  • Page 121: Optics

    Optics 13.1 Spectrometer adjustments In order to adjust devices with auto adjustment of spectrometer use a Skantest_lt program. Operator has choice use a three mode to adjust spectrometer. Two automatic and one manual. [FAST] mode – this mode quickly adjust the spectrometer. Is usefully in case when band height drops under the influence of temperature.
  • Page 122 Manual mode. Operator can calibrate the spectrometer semi manually. Arrows identify the screws. Rules of alignment are that same like crossing the spectrometer (see: crossing the spectrometer. SOCT Copernicus PLUS - Service Manual...
  • Page 123: No Signal In Skantest Program

    13.2 No signal in skantest program 1. Turn on program skantest, run Scan + Mov and Sled then go to bookmark Sled. If there is no signal, as it is shown below, make measurement of power in objective lane ( see: Checking and adjusting power on objective lane).
  • Page 124 4. Check the power on objective lane if the power is incorrect arrange correct value/ the right value. ( see: Checking and adjusting power on objective lane). Even if you arrange correct value of power, peak is still too high you can lower its height using filter on reference lane ( see: Regulation of filter on reference lane).
  • Page 125: Defocusing Of Spectrometer (Too Wide Peak)

    13.3 Defocusing of spectrometer (too wide peak) 1. When you get the peak like in the picture below after using function Up_Dn or Dn_Up: 2. Open SOCT cover ( see: Opening cover), then turning correct screws of the trolley web AVIIVA, set peak.
  • Page 126 3. Turn the right screw in accordance with clockwise if the peak is getting wider turn the screw in the opposite direction ( make turnover of 180 ) and then turn the screw from the left side in the same way. Turn alternately until you will get the highest peak ( always turn each of screw of 180 4.
  • Page 127 5. Using commands Full calibrate spectrometer. The picture below shows window’s program after inclusion of function Full. 6. When you finish calibration by using function Full and the height of spectrum does not fit in range 3100-3400, use function Fast. If the height of spectrum still does not fit in the correct range correct it using filter reference line ( see: Regulation of filter on reference line).
  • Page 128: Too Low Spectrum On Skantest

    13.4 Too low spectrum on skantest 1. When the spectrum on skantest is too low after turning on and using function Fast. 2. Measure the power from the objective lane ( see: Checking and adjusting power on objective lane). When the power is incorrect arrange it to the normal value. 3.
  • Page 129 4. If the height of spectrum does not fit in the correct range arrange it using filter of reference line Regulation of filter on reference line. 5. If spectrum’s height achieved correct value turn off program skantest. 6. Turn on program Copernicus and perform control scan. 7.
  • Page 130: Too High Spectrum On Skantest

    13.5 Too high spectrum on skantest 1. When the spectrum on skantest is too high after turning on and using function Fast. 2. Measure the power from the objective lane (see: Checking and adjusting power on objective line). When the power is incorrect arrange it to the normal value. 3.
  • Page 131: Checking And Adjusting Power On Objective Lane

    13.6 Checking and adjusting power on objective lane Tools needed to do activities shown in this instruction: Power meter; program skantest; hexagonal keys 1,5; 2. 1. Prepare the power meter. 2. Set wavelength at 840nm. Head of Power meter 3. The head of Power meter install on objective lens of SPOCT Copernicus and turn up screw. SOCT Copernicus PLUS - Service Manual...
  • Page 132 4. Run scantest program. 5. Mark scan+Mov and then SLED. 6. Check value of power on the power meter. (Correct value is 950-1020 μW) . 7. if the power is set correctly you can finish measurement. SOCT Copernicus PLUS - Service Manual...
  • Page 133 8. If the power value is incorrect . On the back side of SOCT in left down corner is small tap. Remove it. 9. You see three screws from attenuator. 10. Use a hexagonal tool size 1.5 to rotate first screw about 45 degree, all time you control value on power meter.
  • Page 134 12. When you set up correctly value on objective lane, check the properly settings spectrum in skantest program. If spectrum is correct then go to point 13, if spectrum is incorrect then go to the point 13. Turn off tap, remove head of power meter from objective lens of SPOCT Copernicus, turn off program skantest, turn on program Copernicus and do control scan.
  • Page 135: Setting Fixation Point To The Centre Of The Cross

    13.7 Setting fixation point to the centre of the cross Tools needed to do activities shown in this instruction: - web camera; - program skantest; -viewfinder mounted on objective lens; - hexagonal key 2. 1. Run skantest program. 2. Mark scan+Mov and then SLED. When you switch on SLED, don’t look into objective lens!!! SOCT Copernicus PLUS - Service Manual...
  • Page 136 3. Mark Scanner Ctrl lap. 4. If you moved scroll scanner X or scanner Y press Center button . 5. Select Cross in menu ‘Program control’ and press a button Upload. SOCT Copernicus PLUS - Service Manual...
  • Page 137 6. Press Run button. You will hear how scanners work. Now you can look at objective lens. SOCT Copernicus PLUS - Service Manual...
  • Page 138 7. Go to Digital I/0 lap. Select Center in Fixation menu. 8. Look into objective lens. You will see red cross and green fixation point. If the setup is correct, fixation point is in the centre cross. If the setup is correct don’t read this instruction, if is incorrect go to the next point.
  • Page 139 10. Press button Stop in bookmark Scanner ctrl in program skantest and the next press button Centre at scroll scanner X and scanner Y. 11. Set up viewfinder on objective lens. SOCT Copernicus PLUS - Service Manual...
  • Page 140 12. Using a webcam check if out coming light beam falls onto the centre of the calibration cross. 13. If the beam does not fall in the centre, remove upper part of the cover (see: Opening cover). If beam falls in the centre of the cross, proceed to the point Dichroic mirror SOCT Copernicus PLUS - Service Manual...
  • Page 141 14. On the holder of diachronic mirror you will find three screws. 15. By screwing-in and screwing-out consecutive screws ( observing image from the webcam At the same time) set the beam in the right position. Rotating screw number 1 clockwise causes shifting the beam down and to the right.
  • Page 142 16. After setting the beam in the centre of the cross, remove viewfinder. 17. If fixation point is in the centre of the cross then proceed to the point 18. If beam is centered, and fixation is moved then fixation position needs to be set by regulating of the mirror.
  • Page 143 22. If fixation point is moved left/right then detaching mirror will be needed. 23. While mirror is detached, add or take off some pads in order to set fixation in proper point. Adding pads will move fixation to the left and taking them off will move it to the right. 24.
  • Page 144: Checking And Adjusting The Polarization

    13.8 Checking and adjusting the polarization Tools needed to do activities shown in this instruction: -program skantest; -artificial eye; - spanner size 5.5 -hexagonal key size 2 1. Open the SOCT cover (see: Opening cover 2. Run skantest program. 3. Mark scan+Mov and then SLED. and go to the SLED tab. 4.
  • Page 145 5. Adjust spectrum at 1000 value (see: Regulation of filter on reference lane). 6. Install artificial eye on objective lens. 7. Cover the signal from reference line. SOCT Copernicus PLUS - Service Manual...
  • Page 146 8. Rotate screws on artificial eye. Set the band- height on highest value. During rotating observe the spectrum. 9. Rotate screws 1 on artificial eye. Set the band-height on value around 1000. SOCT Copernicus PLUS - Service Manual...
  • Page 147 10. Uncover signal from reference lane. See interference pattern. SOCT Copernicus PLUS - Service Manual...
  • Page 148 11. You can move refraction to obtain proper density of spectrum. SOCT Copernicus PLUS - Service Manual...
  • Page 149 12. Perform adjustments of the optical fibre holders, to achieve spectrum like shown on image. 13. Good interference spectrum: 14. Uninstall artificial eye from objective lane. 15. Set the band height to 3100-3400 level (see: Regulation of filter on reference line).
  • Page 150: Checking, Adjusting And Exchange Of Scanners

    13.9 Checking, adjusting and exchange of scanners Tools needed to do activities shown in this instruction: -Power meter; - IR Detector Card; -program skantest; - viewfinder on objective; - hexagonal keys; 0,9; 2; 2,5. 1. Turn on Copernicus and run scantest. 2.
  • Page 151 SOCT Copernicus PLUS - Service Manual...
  • Page 152: Exchanging Of Scanner Y

    13.9.1 Exchanging of scanner Y 1. If scanner Y is broken, disconnect its plug. 2. Undo the screw. SOCT Copernicus PLUS - Service Manual...
  • Page 153 3. Extract the scanner. 4. Look for broken mirror. 5. Detach rest of broken mirror from holder . 6. Install new mirror. Mirror holder screws SOCT Copernicus PLUS - Service Manual...
  • Page 154 7. Install scanner in scanner holder. 8. Install viewfinder on objective lens. 9. Install webcam. SOCT Copernicus PLUS - Service Manual...
  • Page 155 10. Check if there is no collision between scanners. 11. Turn device on and the next switch on program skantest. 12. Referring to the camera image set the beam position in centre by rotating Y scanner. Rotation of Y scanner causes in horizontal movement of the beam. SOCT Copernicus PLUS - Service Manual...
  • Page 156 13. Screw a bolt. 14. Switch off program skantest and check collision between scanners. 15. Close the SPOCT head (see: Closing cover). SOCT Copernicus PLUS - Service Manual...
  • Page 157: Exchanging Scanner X

    13.9.2 Exchanging scanner X 1. If scanner X is broken down, disconnect its plug. 2. Unscrew bolt . SOCT Copernicus PLUS - Service Manual...
  • Page 158 3. Extract scanner. SOCT Copernicus PLUS - Service Manual...
  • Page 159 4. Looking for broken down mirror. 5. Unscrew rest broken down mirror . 6. Install new mirror. Mirror holder screws 7. Install scanner in scanner holder. SOCT Copernicus PLUS - Service Manual...
  • Page 160 8. Install viewfinder on objective lens. 9. Install webcam. 10. Check if do not have collision between scanners. 11. Switch on device and program skantest. SOCT Copernicus PLUS - Service Manual...
  • Page 161 12. Referring to the camera image set the beam position in centre by rotating X scanner. Rotation of X scanner causes in horizontal movement of the beam. 13. Screw Bolt. SOCT Copernicus PLUS - Service Manual...
  • Page 162 14. Switch off program skantest and check collision between scanners. 15. Close SPOCT head. SOCT Copernicus PLUS - Service Manual...
  • Page 163: Correcting Scanner Setup

    13.9.3 Correcting scanner setup. 1. If scanners are not destroyed mechanically and there is no signal on lens course check if there is a beam outgoing from collimator. 2. To check if there is beam outgoing from collimator, use IR sensitive paper by placing it behind collimator.
  • Page 164 10. Rotate the scanner all time monitoring view on the webcam. 11. Set beam in centre position and tighten the screw. 12. Check if there is no collision between scanners. 13. Close the cover (don’t follow this step if you have to adjust Y scanner too)(see: Closing cover).
  • Page 165 14. Loosen the scanner Y screw . 15. Rotate the scanner all time monitoring view on the webcam. 16. Set beam in centre position and tighten the screw. SOCT Copernicus PLUS - Service Manual...
  • Page 166 17. Check if there is no collision between scanners. 18. Close the cover (see: Closing cover). SOCT Copernicus PLUS - Service Manual...
  • Page 167: Cleaning Optics

    13.10 Cleaning optics Tools needed to do activities shown in this instruction: - cotton cloth; - brush; - - optical fiber cleaner 1. Clean objective lens with clean soft cloth. a) From outside. b) Internal surface. Move the refraction on maximum plus position. Open SPOCT head cover (see: Opening cover).
  • Page 168 - check the continuity of optical fibres does mean are there the visible cut or crease on cover. 4. Cleaning the optics: We make cleaning lens using designed to this especially tools, in dependence from requirements: - dust - remove by blowing surfaces with compressed air from converter, alternatively using brush with soft natural hair.
  • Page 169 Then scanner X Same steps should be performed why cleaning optics above the scanners. 6. Cleaning the fibre optics. Collimator on objective lane SOCT Copernicus PLUS - Service Manual...
  • Page 170 Fiber cleaner 7. Close SPOCT cover (see: Closing cover). SOCT Copernicus PLUS - Service Manual...
  • Page 171: Regulation Of Filter On Reference Lane

    13.11 Regulation of filter on reference lane Tools needed to do activities shown in this instruction: webcam; program skantest;; two spanners size 5,5 . 1. Open the cover (see: Opening cover). 2. Prepare optic cleaning set, two spanners size 5.5. 3.
  • Page 172 5. Switch on Scan+Mov and SLED. 6. Switch on overlap SLED. SOCT Copernicus PLUS - Service Manual...
  • Page 173 7. Take the special rag for cleaning optics. Carefully try rotating a filter just a little. (if you cannot do it gently resistance is too big go to the next point). Observe the spectrum and set value on 3200-3500. Spectrum of graph appears only when reference line is clear. 8.
  • Page 174: Checking Of Sled Output Power, Checking Attenuator

    13.12 Checking of SLED output power, checking attenuator Tools needed to do activities shown in this instruction: scantest program; Power meter . 1. Open the cover (see: Opening cover). 2. Equip power meter with detector for measuring output power from optic fibre. 3.
  • Page 175 4. Run Skantest program. 5. Switch on Scan+Mov and Sled. 6. Check Reading from power meter. Accurate value should exceed 3,5 7. If output Power is incorrect check if diode powering is accurate. 8. If Power is correct, reconnect dumper’s plug. 9.
  • Page 176 12. Measure output power from sample course and signal from reference course. If measured values are correct go to the next point. If there is no signal on one or both courses it appears that beam splitter is broken. Unit should be sent back to the Optopol. SOCT Copernicus PLUS - Service Manual...
  • Page 177 13. Exchange the dumper. To exchange attenuator unscrew two screws shown on the image. 14. Exchange old dumper with new one. Attach diode and beam splitter connectors. 15. Set Power on the sample course (see: Checking SLED output power, checking attenuator).
  • Page 178: Crossing The Spectrometer

    13.13 Crossing the spectrometer. 1. Turn on program skantest, run Scan + Mov and Sled then go to bookmark Sled. Use function Up_Dn or Dn_Up. To Change camera trolley position and cross spectrometer. To change the position of trolley you can also use spectrometer manual control for left and right screw.
  • Page 179: Spectrometer Stability Test

    13.14 Spectrometer stability test 1. Turn on the computer and SOCT. 2. Run skantest program and mark scan+mov and SLED. 3. After 30 minutes perform FULL spectrometer adjusting by pressing [FULL] tab. 4. After calibration note the spectrum height and outgoing power on the objective line. 5.
  • Page 180: Faq

    14. FAQ Unit functioning in skantest program is proper, but in SOCT Copernicus PLUS software there is no tomogram, There is problem with software or synchro cable. Check or reinstall software or replace cable and perform tests. Power measured on the exit of the diode is lower than 3,5mW. If diode have connectors between optic fibre outgoing from diode and coupler, the power loss on the connectors is too high.
  • Page 181 While testing power stability, power fluctuation extends allowed limits. Unstable attenuator functioning. Readjust attenuator again (following attenuator adjusting rules) and after one hour of cooling perform test again. If that doesn’t help exchange attenuator with new one and after cooling the unit perform test again. Scan’s saturation is low.
  • Page 182 Q10: While performing scans, it’s able to notice that on height about 700 scan is almost invisible. A10: Perform spectrometer stability test, check CL cables. Bad scan Good scan Q11: Horizontal line appears on tomogram on height about 500. A11: Beamsplitter optic fiber connectors are dirty.
  • Page 183 Q14: Horizontal line appears on tomogram on height about 50 after 30 minutes disperses to the trail. A14: Line comes from the camera. If after camera have warmed (about 30 min.) it disperses and is barely visible on the camera, we leave it and do nothing. If it’s strongly visible, we replace the camera.
  • Page 184 Q30: After centering scanners scanner position is wrong. A30: Broken scanner controller D/A converter - exchange converter Wrongly attached scanner steering wiring – check correctness of connection or exchange wiring Broken CPU board 151-506 – replace CPU board Q31: While normal functioning device goes to the “base” position. A31: Check correctness of attachment of all plugs on PSB board and motor control board Broken PSB board or Movement control board 151-517 –...
  • Page 185 Q34: Noise on tomogram. A34: Oversaturated spectrum. Adjust gradient filter on reference lane. __________________________________________________________________________________ Q35: Cannot detect recognizable video source. A35: CL cable disconnected. No voltage on AVIIVA camera. __________________________________________________________________________________ Q36: No tomogram preview. A36: CL cable disconnected. No voltage on AVIIVA camera. Wrong camera is selected in NI drivers.
  • Page 186 __________________________________________________________________________________ Q37: SOCT_26.exe – error. A37: Synchronization cable disconnected. CL cable disconnected (after cable connection device must be restarted). No voltage on AVIIVA camera (after cable connection device must be restarted). ________________________________________________________________________________ Q38: Connection with SOCT error. Please Reconnect. A38: No voltage on AVIIVA camera (during work).
  • Page 187 ________________________________________________________________________________ Q41: Calibration fail. Please run scantest_lt. A41: NI driver is not installed. ________________________________________________________________________________ Q42: Unable to start acquisition module. A42: NI driver is not installed. ________________________________________________________________________________ Q43: The serial read did not complete with the specified timeout period. A43: Wrong camera selected in NI drivers.
  • Page 188: Spare Parts List

    Spare parts list Part Name Image Comment number Computer with PCI-1428 and Video frame grabber, 1. 4 fully installed A001P and configured, Keyboard, Mouse, Mouse Camera Link Frame Grabber 2. 7 A004P card PCI-1428 3. 8 Video frame A005P Grabber card 4.
  • Page 189 Video/synchro cable B002P 1.8m long B003P Power supply 8. 1 cable 1,8m long B004P Camera link cable Type depend from Power Supply 230V Separation or 115V Transformer B005P Transport protection C001P handle 12. 1 Chinrest C002P adapter SOCT Copernicus PLUS - Service Manual...
  • Page 190 Single use 13. 1 chinrest labels C003P Chinrest pin C004P 15. 1 Dust cover C005P 16. Head cover flap C006P SOCT SOCT Transportation Transportation D001P box (full) box (full) SOCT 18. 1 transportation D002P (top cover only) SOCT Copernicus PLUS - Service Manual...
  • Page 191 SOCT 19. 1 Transportation D003P (base only) X movement X movement D004P toothed bar toothed bar Y movement Y movement D005P toothed bar toothed bar Z movement Z movement D006P toothed bar toothed bar Movement Movement engine D007P engine rack rack Metal brackets for SOCT...
  • Page 192 Movement 27. 2 control board E003P (151-517) Sled diode E004P Board 151-512 E005P (CPU Slave) E006P Changed to 151-126 30. 2 USB connection module, compatible board 151-125 Spectrometer DEPENDS FROM 31. 2 control board E007P SERIAL NUMBER (151-539) Spectrometer DEPENDS FROM control board E007AP SERIAL NUMBER...
  • Page 193 34. 2 Forehead E009P frame 35. 2 Servomechanis E010P m of refraction 36. 3 Servomechanis E011P m of preview E013P Stepper engine DEPENDS FROM Motor for SERIALN NUMBER From S/N: 154229 E014P chinrest is a new type module Scanner set X and Y 39.
  • Page 194 Scanner set Y 41. 3 (Scanner + E017YP Cable + Controller ) 42. Scanner X cable E018XP 43. Scanner Y cable E018YP 44. 3 X scanner E019P mirror 45. 3 Y scanner E020P mirror Micro switches with screws for E021P chinrest 47.
  • Page 195 TYPE DEPENDS FROM SERIAL Spectrometer E024P NUMBER servomechanis m module TYPE DEPENDS Spectrometer FROM SERIAL servomechanis E024AP NUMBER m module Scanners converter E025P board Schulter filter E026P Transitional E028P board 151-525 Transitional E029P board 151-526 Transitional E030P board 151-527 SOCT Copernicus PLUS - Service Manual...
  • Page 196 Main lens F001P 56. Gradient filter F002P 57. 4 F003P Attenuator 58. 4 F004P Lens cover Spare motor F005P belt for chinrest with set of screws SOCT Copernicus PLUS - Service Manual...
  • Page 197 F008P head frame to main frame Screws to fix Set (10 screws and F009P ground line to 20 nuts) the headrest Up to date part pricelist can be downloaded from OPTOPOL ftp server http://dane.optopol.com/ SOCT Copernicus PLUS - Service Manual...
  • Page 198: Service Tools

    Service tools Name Part number Image Long hexagonal tolls Size: 2, 2.5, 4 Small hexagonal set Size: 2, 1.5, 1.3, 0.9, 0.7 Hexagonal set Size: 1.5, 2, 2.5, 3, 4, 5, 6, 8, 10 Sparks Size: 2 x 5.5 Calibrating sight (viewfinder) USB digital camera on support...
  • Page 199 Artificial eye Power meter Thorlabs PM120 Optic fibre adapter + lens adapter for set PM120 Infra-red detector Cleaning set: Cloth, air pump with brush, disposal cleaning paper, Grease Kluber Lubrication Kluberpaste UH1 84-201 100ml Grease Kluber Lubrication Klubersynth RA-3502 100ml SOCT Copernicus PLUS - Service Manual...
  • Page 200 Grease Naftochem Smar ŁT-43 100ml Tool for replace main lens Optic fiber connector cleaner CLETOP-S Programming jumper hexagonal key for mirror exchange SOCT Copernicus PLUS - Service Manual...

Table of Contents