The sterilizer periodically reminds the user about recommended "routine" maintenance operations that must
be carried out in order to ensure the proper operation of the device. The reminder notices are displayed on
the screen when a sterilization cycle is started.
The frequencies indicated have been calculated for "standard use" of the device, used correctly and
positioned in a suitable environment. Should you notice a significant lose in performance, repeated
occurrence of alarms or visible degradation of the parts subject to wear, it is advisable to carry out the
maintenance operations earlier than the due dates programmed in the system.
Daily
Clean the door gasket and plate with a clean cotton cloth soaked in a weak solution of
water and vinegar. Dry the surfaces and remove any residue before using the device.
Clean external surfaces using a clean cotton cloth dampened with water and, if needed a
neutral detergent.
Weekly
Clean the sterilization chamber and relative accessories.
Disinfect external surfaces.
Monthly
Clean the internal distilled water tank.
Safety valve maintenance.
Clean (or replace) the drain filter.
Annual or every
Replace the door gasket.
1000 cycles
Every 3 Years or
Recommended complete maintenance of the sterilizer by an authorized dealer.
3000 cycles
Bravo Maintenance
Display Warning Messages
Warning Message
Chamber Filter Cleaning
Bacteriological Filter Replacement Every 400 cycles
Door Gasket Replacement
General Service
Bravo Maintenance
Frequency
Every 200 cycles
Every 1000 cycles
Every 3000 cycles