Table 8-1 Coarse Zero Adjustment; Table 8-2 Binary Count Sequence For Coarse Zero Jumper Pattern - Honeywell 7866 Operation And Maintenance Manual

Triple range digital h2 and co2 analyzer/indicator and gas sampling panel
Table of Contents

Advertisement

Troubleshooting
Coarse Zero Adjustment (R11 and R16)
It may be necessary at some time to return the control-unit 20-turn fine ZERO (R16) to its center-of-travel
position. To do so requires a jumper-wire change in the sensing unit.
Step
Open the sensing unit and observe the jumper pattern used at the top of the round circuit card
1
(terminal board at bottom). Note that each pin provided for the six jumpers is identified with a 1
or a 0. Determine the binary number formed by the jumper positions used and record this
number. The letters identify the color of the corresponding jumper.
2
If the analyzer output current or voltage is above zero or other low-end value and the ZERO
adjuster is at its counterclockwise limit, determine that binary number which is one digit lower
than the number recorded and change the necessary jumper positions to form this number.
For example, if the jumpers are positioned to form binary number 111001, then the jumpers
must be repositioned to form the next lower number binary 111000. Change the sixth jumper
from 1 to 0.
If the analyzer reading is below the low-end range value and the ZERO is at its clockwise
limit, the zero must be shifted upscale. Position the jumpers to form the next higher number,
e.g., from binary 011010 to 011011. Move the right-end jumper from 0 to 1.
Refer Table 8-2 for the sequence used in binary counting. The card is screened with MSB
(Most Significant Bit) and LSB (Least Significant Bit).
The above jumper change should permit the return of fine ZERO adjuster R16 to its center-of-
3
travel position and allow completion of the analyzer zero adjustment. If it does not, simply
increase or decrease the binary number by one digit more. A change of one binary count is
equivalent to one-half travel of control unit ZERO (R16). If jumpers are changed inadvertently
before the binary number has been recorded (or if thermistor detectors have been replaced)
re-establish the correct jumper pattern. Measure the sensing-unit current output by connecting
a meter in series with terminal 12 at the sensing unit while zero-gas flows through the sensing
unit. Proceed through the steps in Table 8-3 in numerical sequence only until the current
output measures between 250 µA and 240 µA. Jumpers that have been placed in their final
positions, not been moved, are marked X. Start with binary 32 (100000).

Table 8-2 Binary Count Sequence for Coarse Zero Jumper Pattern

MSB
LSB
0
000000
1
000001
2
000010
3
000011
4
000100
5
000101
6
000110
7
000111
8
001000
9
001001
68
7866 Triple Range Digital H

Table 8-1 Coarse Zero Adjustment

MSB
LSB
16
010000
17
010001
18
010010
19
010011
20
010100
21
010101
22
010110
23
010111
24
011000
25
011001
and CO
Analyzer/Indicator - Operation and Maintenance
2
2
Action
MSB
LSB
32
100000
33
100001
34
100010
35
100011
36
100100
37
100101
38
100110
39
100111
40
101000
41
101001
MSB
LSB
48
110000
49
110001
50
110010
51
110011
52
110100
53
110101
54
110110
55
110111
56
111000
57
111001
9/06

Advertisement

Table of Contents
loading

Table of Contents