This controller cannot receive Emails, so it is suggested that you configure the "From Email:" window with
a non-Email style address that will make it easy for you to determine which controller sent the Email. For
Email technical reasons, the entry in the "From Email:" window cannot have spaces. See Figure 4-2.
If you do not know your SMTP IP Address for outgoing Email, then contact your Information
Technologies (IT) representative. If your PC is on the same LAN that will be used by the controller and
which also connects to the Email server, then the SMTP IP Address may generally be found by opening a
DOS shell and typing:
ping smtp.[your domain name and extension, i.e., "yourisp.com"]
The content of the Emails sent by this controller contains the Alarm that triggered the Email, its settings
and the current value (if applicable) of the monitored variable. For example, the content of an Email
triggered by Alarm 1 Setpoint 1 that is configured to monitor Input 1 would look something like this:
Name: Alarm 1 SP1, Type: INPUT1, Event: HIGH/END, Value = 500.00, Actual = 712.69
ATTENTION
Instruments will always send Email time-stamped with the date that the Ethernet Software in the
instrument was last modified.
If the SMTP address on your network is changed, such as can happen when a server is replaced, then
you must reconfigure the Email SMTP IP address in this instrument to match.
9/06