Permissible Wiring Bundling
Bundle No.
1
2
3
3.7

Wiring Between Sensing Unit, Controller, and Indicator

Wiring Connections at the Sensing Unit
Step
After the sensing unit is mounted, carefully remove the top threaded cover by turning it
1
counterclockwise. (Loosen the set screw.)
2
Two openings are provided for cable entrance; make certain the plug screw is tightened firmly into
the unused right-hand opening.
3
Use 1/2 NPT threaded electrical conduit, and mount the special adapter fitting supplied with the
analyzer.
4
Wire as shown in Figure 3-5.
EXPLOSION-PROOF SENSING UNIT REQUIREMENTS
Always use conduit for electrical wiring and follow special installation procedures.
Make certain fully qualified personnel make this installation and all local regulations are observed.
Failure to comply with these instructions could result in fire caused by explosion inside the housing.
9/06
7866 Triple Range Digital H

Table 3-2 Permissible Wiring Bundling

• Line power wiring
• Earth ground wiring
• Line voltage control relay output wiring
• Line voltage alarm wiring
Analog signal wire, such as:
• Input signal wire (thermocouple, 4 to 20 mA, etc.)
• 4-20 mA output signal wiring
Digital input signals
• Low voltage alarm relay output wiring
• Low voltage wiring to solid state type control circuits
• Low voltage wiring to open collector type control circuits
Action
and CO
Analyzer/Indicator - Operation and Maintenance
2
2
Wire Functions
Installation
27