Angle display
Hide thumbs Also See for LH70:

Advertisement

3-2. Setting operation

First of all, in the basic settings, set the measurement unit resolution to the angular resolution that takes into account the output pulse
per rotation of the rotary encoder. For details on the basic settings, refer to the separate initial setup manual of display unit.
Next, set the display resolution and scaling value in the Advanced settings. Unless the number of rotary encoder output pulses is the
same as the number of pulses converted to minutes and seconds, it is always necessary to correct the number of pulses by scaling.
(Factory default scaling: 1.000000)
Select the measurement unit resolution that can be used in the basic settings. Check the number of
Basic Settings
output pulses per rotation of the rotary encoder, select the angular resolution that can be set from the
table below, and set it.
SIG RES
Measuring unit resolution
Rotary encoder output pulse number and configurable angular resolution
Number of output pulses
Classification
(N / 1 rotation)
216 ≤ N < 2,160
2,160 ≤ N < 21,600
21,600 ≤ N < 129,600
129,600 ≤ N < 1,296,000
1,296,000 ≤ N
Advanced Settings
Set to advanced settings mode
Determine resolution
How to calculate the scaling magnification
Magnification =
Encoder output pulse number (1 rotation)
Settable resolution
10min.
10min. , 1min.
1min. , 10sec.
10sec. , 1sec.
1sec.
Enter the advanced settings mode and set the display resolution and scaling.
Set the display resolution
Press the key to display the scaling settings
SCALING
Enter the scaling magnification using the numeric keys
P
LH70_LY70 Angle display Quick reference manual
10min.
1min.
10sec.
1sec.
Selection of display resolution
10min.
10min.
1min.
10sec.
1sec.
Press the key of
the axis to set
Scaling
magnification input
magnification
P
2,160
21,600
129,600
1,296,000
Press the key
several times to
select the resolution
or
1min.
or
10sec.
1sec.
or
Select the same resolution as the basic setting
10min. 1min. 10sec. 1sec.
* The resolution of the basic setting is
selected as the initial value.
Return to normal display
Save input value
default
Measurement unit resolution set
in the basic settings
10min.
1min.
10sec.
1sec.
6/6