2-3. How to install

A typical example for pasting scales in parallel is shown below.
SL130 and SL331 have a strong scale base material, so if they are
attached to an arc surface, the scale edge may bounce off. It is
recommended to fix the base material of the scale with a suitable
clamp plate.
For more information on pasting scales, refer to the head unit
instruction manual.
Typical example of how to install
(1) Set up parallel pins on the flat
surface and attach the scale. Remove
the parallel pin after installation.
Scale
Scale cross section and clearance
SL130
Reader head
Protective ribbon
Scale base
Width: 10mm

2-4. Setting operation

First of all, in the basic settings, set the length measurement unit resolution to the output resolution of the connected digiruler. For
details on the basic settings, refer to the separate initial setup manual of display unit. Then enter the display resolution and mounting
radius values in the advanced settings. Compare the mounting radius value with an angle reference device (right angle ruler: 90
degrees), etc., and narrow down the radius value so that the angle error is minimized.
In the basic settings, set the output resolution of the connected digital ruler.
Basic Settings
It is not angular resolution.
SIG RES
Measuring unit resolution
Advanced Settings
Set to advanced settings mode
Press the key to set the resolution you want to display
10min.
*: If the radius cannot be entered and an error occurs, the angle display for the measurement unit is too small, so increase the radius or
coarsen the resolution of the angle display.
(2) Make a stop surface for mounting
the scale. Consider the thickness of
the scale and the amount of clearance
with the reader head
Parallel pins
Scale
Reader head
Clearance
Protective ribbon
0.8mm
Scale
Scale base
Width: 9mm
Width: 10mm
Set to the advanced setting mode and enter the display resolution and radius in the display resolution
setting.
Set the display resolution
1min.
10sec
1sec
LH70_LY70 Angle display Quick reference manual
Clamp plate to prevent bounce at
the scale edge of SL130, SL331
(3) Provide a mounting groove with
a width slightly larger than the scale.
Step for stop surface
SL331
Clearance
0.5mm
Clearance
Scale
0.35±0.1mm
Width: 9mm
0.5μm
1μm
2μm
20μm
25μm
50μm
Press the key of
the axis to set
Enter the radius *
Determine resolution
Enter the radius
with the numeric
keys(unit: mm)
Example of jump prevention
Clamp plate
Scale
Clamp plate
Scale
Mounting groove
SL700
25mm
Reader head
PL101
Scale
9mm
Thickness: 0.3mm Width:9mm
5μm
10μm
Press the key
several times to
select the resolution
100μm
Set the angle of the digiruler
For LH counter
For LY counter
Return to
Determine
normal display
the radius
Screw
4/6