General Cleaning - Bayer HealthCare CLINITEK 50 User Manual

Table of Contents

Advertisement

Wet a cotton-tipped stick
3
with the NaOH. Thoroughly
clean the trough and side areas
of the table until all urine is
removed.
Rinse the table thoroughly
4
with water to remove the NaOH.
Wipe the surface of the table
5
and the grooves on the under-
side of the table with a damp
cloth. Then dry the table with a
soft cloth or tissue and insert it
into the instrument.

General Cleaning

Always keep the outside of the
CLINITEK
50 instrument clean
®
and free of dust. You can wipe
the outside with a damp (not
wet) cloth and a mild detergent.
Wipe the display screen with
a soft, nonabrasive cloth that
has been dampened with a mild
glass cleaner. Do not spray the
glass cleaner directly onto the
screen. Do not use laboratory
wipes, such as Kimwipes
they may scratch the screen.
bar with the NaOH!
solvent, oil, grease, silicone
spray, or lubrication on the
instrument.
The keypad and display may
be disinfected using the same
solutions as for the test strip
table (see page 19). Wipe the
solution on and allow to remain
for 10 minutes. Rinse using a
clean cloth dampened with
®
, since
water, then dry.
Do not touch the white
Do not use any type of
21