Specifications

Power Required:
Power Transformer:
Input: 100-250 V~, 50/60 Hz,
0.5-0.3 A, grounded
Output: +9 V
Optional Battery Pack —
Holds 6 "AA" standard alkaline or
rechargeable NiCad batteries
Up to 200 tests can be run when
using new alkaline batteries,
fewer with rechargeable batteries.
Instrument Fuse:
4A, 63V, Quick-Acting
( not replaceable by the user)
Line Leakage Current:
<0.5 milliamperes in normal
condition
<3.5 milliamperes in single
fault condition
( Testing protocol and allowable limits as
specified by the safety standards for
laboratory equipment outlined in UL 3101-1,
CSA 22.2, No. 1010.1, and IEC 1010-1.)
Computer Interface:
Customized RJ11 Cable
( For use in connecting to a
computer only)
Dimensions:
Depth — 23.5 cm (9.2 in.)
Width — 15.2 cm (6.0 in.)
Height — 15.5 cm (6.1 in.)
Weight:
Instrument only — 1.25 kg (2.8 lbs.)
Instrument with Battery Pack —
1.42 kg (3.1 lbs.)
28
, 2.78 A
Ambient Operating Temperature
Range:
18°C to 30°C (64°F to 86°F)
Optimum Operating Temperature
Range:
22°C to 26°C (72°F to 79°F)
Ambient Operating Humidity Range:
20% to 85% Relative Humidity
Safety Standards:
The CLINITEK
®
50 Urine Chemistry
Analyzer (Models 6510 and 6510A)
conforms to the Low Voltage Safety
Directive 73/23/EEC. The instrument
is listed by the Underwriters'
Laboratories, Inc. (UL) and the
Canadian Standards Association
(CSA) as certified and complies
with the safety standards specified
in UL 3101 and CSA-C22.2, No. 1010.1.
Model 6510A conforms to EMC
Directive 89/336 Amendment
92/31/EEC.
Warning:
This instrument must be used in the
manner specified in this User's Guide
in order to provide the safety and
performance standards specified.
Symbols:
Refer to User's Guide for
complete information on
using this instrument.
DC Voltage
Power Input
Computer Interface
CLIA Status:
The CLINITEK 50 Analyzer is "Waived"
per CLIA '88.