Commands - A&D BM-5 Instruction Manual

Analytical balance
Table of Contents

Advertisement

Commands

16.3.
Command List
16.3.1.
Note Before transmitting a command, add the terminator that is specified in the "Terminator
(Crlf)" parameter of "Serial interface (5if)" in the function table.
Commands to query weighing data
C
Q
S
SI
SIR
Commands to control the balance
CAL
OFF
ON
P
PRT
R
RNG
TR
TST
U
Commands to request data
?ID
?PT
?SN
?TN
Command to set data
PT:.
g
#1
Command R assumes the point as zero and sets the display to zero.
Command TR cancels the tare weight when the displayed value is greater than zero.
The canceled tare value can be read using the command ?PT.
Command PT:.
An example of this command sets a negative target value and fills with the sample until the
display becomes zero. For details, refer to page 94.
Step 1
Place a container.
Step 2
Set the display to zero using the R command.
Step 3
Set a target value using the PT: command.
Cancels the S or SIR command.
Requests the weighing data immediately.
Requests the weighing data when stabilized.
Requests the weighing data immediately.
Requests the weighing data continuously.
Same as the
key.
CAL
Turns the display off.
Turns the display on.
Same as the
ON:OFF
Same as the
PRINT
Same as the
RE-ZERO
Same as the
RANGE
Tares the balance.
Performs calibration test.
Same as the
MODE
Requests the identification number.
Outputs the tare value.
Requests the serial number of the balance.
Requests the model name of the balance.
Sets the tare value.
g sets the tare value digitally.
Calibration using the internal mass.
key.
Turning the display on or off.
key.
key.
Zero display.
key.
Changing the readability.
Canceling the container's weight.
Calibration test using the internal mass.
key.
Changing the unit.
90
#1
#1
#1
#1
BM series