Commands - A&D HR-300i Instruction Manual

Hr-i series analytical balances
Table of Contents

Advertisement

Commands

13.3.
Command List
13.3.1.
Note A command has a terminator added, that is specified using "
function table, and is sent to the balance.
Commands to query weighing data
C
Q
S
SI
SIR
Commands to control the balance
CAL
OFF
ON
P
PRT
R
RNG
TR
U
Commands to request data
?ID
?PT
?SN
?TN
Command to set data
PT:.
g
(Note 1)
R command assumes the point as zero and sets the display to zero.
TR command cancels the tare weight when the displayed value is greater than zero.
The canceled tare value can be read using the ?PT command.
PT: command sets the tare value digitally.
An example of this command sets a negative target value and fills with the sample
until the display becomes zero.
Step 1 Place a container.
Step 2 Set the display to zero using the R command.
Step 3 Set a target value using the PT: command.
For details, refer to page 49
Cancels the S or SIR command.
Requests the weighing data immediately.
Requests the weighing data when stabilized.
Requests the weighing data immediately.
Requests the weighing data continuously.
Same as the
CAL
Turns the display off.
Turns the display on.
Same as the
ON:OFF
Same as the
PRINT
Same as the
RE-ZERO
Same as the
RANGE
Tares the balance (Cancels the container's weight.)
Same as the
MODE
Requests the ID number.
Outputs the tare value.
Requests the serial number of the balance.
Requests the model name of the balance.
Sets the tare value.
key.
key
key
(Note 1)
key
key
key
(Note 1)
(Note 1)
44
" of the
5if Crlf
(Note 1)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hr-202i

Table of Contents