18-1 Command List

Commands to query weighing data
C
Q
S
SI
SIR
E
P
S C
Note: The "Q" and "SI" commands, the "S" and "
Commands to control the balance
CAL
MCL
MD:nnn
OFF
ON
P
PT:∗∗∗. ∗∗
g
?PT
R
SMP
U
T
TR
Z
E
T
S C
Note: The "R", "T", "Z" and "
Commands to control the memory function
?MA
?MQnnn
?MX
MCL
MD:nnn
UN:mm
?UN
UW:∗∗∗. ∗∗
g
?UW
?ID
?SN
?TN

18. COMMANDS

Cancels the S or SIR command.
Requests the weighing data immediately.
Requests the weighing data when stabilized.
Requests the weighing data immediately.
Requests the weighing data continuously.
Requests the weighing data when stabilized.
Same as the CAL key.
Deletes all data in memory.
Deletes weighing data with the data number nnn.
Turns the display off.
Turns the display on.
Same as the ON:OFF key
Sets the tare value.
The unit added is the unit that is output in the A&D standard format. For
the counting or percent mode, gram is used. Set the tare value so that
the total value of the sample and the tare does not exceed the weighing
capacity. Negative values can not be used.
Requests the tare value.
Outputs the tare value set by the TR or PT: command.
Same as the RE-ZERO key
Same as the SAMPLE key.
Same as the MODE key
Same as the RE-ZERO key
Tares the balance (Cancels the container's weight.)
Available only when the load on the weighing pan is greater than the
zero point. The zero point is set by the R, T or Z command, by pressing
the RE-ZERO key, or using the external input.
Same as the RE-ZERO key
Same as the RE-ZERO key
E
T" commands behave the same.
S C
Outputs all data in memory.
Outputs weighing data with the data number nnn.
Outputs the number of weighing data in memory.
Deletes all data in memory.
Deletes weighing data with the data number nnn.
Changes the unit mass stored in memory with the number of mm (01-20).
Outputs the unit mass number of the selected unit mass.
Sets the unit mass value. e.g.,UW:+0.123
0.123 g: represents a space.)
Outputs the unit mass value of the selected unit mass number.
Requests the ID number.
Requests the serial number.
Requests the model name.
E
P" commands behave the same.
S C
g (to set the unit mass to