Appendix
A.3
X-ray Dose Data
The X-ray dose data is extracted from the X-ray Dose Test Report for the EzRay Air
Wall (VEX-S300W). The IEC collateral standards have measured the X-ray doses of
EzRay Air Wall (VEX-S300W) in the test report. The EzRay Air Wall (VEX-S300W)
has been designed by Part 1. General Requirements for Safety, IEC 60601-1-3.
1. X-ray Dose Table
Instrument
Multi-Dose Meter
Dose Area Product (DAP) Table (65 kVp, 3.0 mA, SSD 200 mm)
Dose Table
t (s)
0.05
0.08
0.10
0.15
0.20
0.25
0.30
0.45
0.50
60
Model Name
Tube Model Name
Generator Model Name
Loading Factor
Manufacturer
Raysafe
FOV: Ø 6 cm
mGy
0.222
2.63
0.357
4.71
0.443
6.09
0.663
9.55
0.909
13.01
1.101
16.47
1.326
19.93
1.781
30.32
2.212
33.78
Test Condition
DG-S0102V1 (Inverter type)
 65 kV, 3.0 mA
Test Equipment
Model
Unfors Xi mAs
/ Unfors Xi R/F & MAM
FOV : 4 x 3 cm
Dose (mGy.cm
1.12
2.00
2.59
4.06
5.53
6.99
8.45
12.87
14.34
EzRay Air Wall (VEX-S300W) User Manual
VEX-S300W
V1-650304
S/N
163288 / 161834
FOV : 2 x 3 cm
)
2
0.56
1.00
2.29
2.03
2.76
3.50
4.23
6.44
7.17