5. Troubleshooting
Troubleshooting
Problem
Equipment is not
turned on.
Control Panel is
not turned on.
No X-ray emission
X-ray emission
works, but
exposure is too
light or
completely white.
X-ray emission
works, but
the exposure is
too dark.
40
Cause
The power switch is not
turned on properly.
Battery discharged
Battery cable is not
connected correctly.
Defective battery
Defective mainboard
Internal cable
disconnected
The generator is cooling.
Defective Remote
Exposure Switch
Internal cable
disconnected
Defective generator
Tube lifecycle termination
The equipment has been
positioned incorrectly.
Exposure time is too
long.
The receptor is facing the
the wrong way.
Exposure time is too
short.
EzRay Air Wall (VEX-S300W) User Manual
Solution
Turn the equipment power switch
off and turn it back on.
Recheck after charging the battery
with a charger.
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Wait for the cooling time (refer to
'Duty Cycle' on page 70).
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Adjust the position of the
equipment.
Decrease the exposure time.
Reposition the receptor.
Increase the exposure time.