Thermo Scientific 700 Series Manual page 42

Hide thumbs Also See for 700 Series:
Table of Contents

Advertisement

Thermo Fisher Scientific
Model
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical
Breaker Requirements
Shipping Weight
Certifications
Declaration of Conformity is available from the factory
Safety Specifications
Indoor Use Only
Altitude - up to 2,000 meters
Temperature - 5°C to 43°C
Humidity - Maximum RH 80% for temperatures up to 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
1
Installation Category II
Pollution Degree 2
2
Class of Equipment I
______________________________________________________________________
1 Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II
which is the category used for instruments in installations supplied from a supply comparable to public mains such as hospital
and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500V for a 230V supply and
1500V for a 120V supply.
2
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity caused
by condensation.
721
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
96.0"W x 40.5" H x 28.8" F-B
(243.8cm x 102.9cm x 73.2cm)
Add 7.9" (20.1cm) to F-B for wall spacer/handle.
66.5"W x 28.0"H x 18.5" F-B
(168.9cm x 71.1cm x 47.6cm)
20.0 cu. ft. (566.3 liters)
Cascade system, (2) hermetically-sealed compressors
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;
1.0" (12.5cm) sub-lids
120VAC, 1 PH, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
20 Amp, 120VAC, Dedicated Circuit,
20 Amp Time Delay Breaker
833 lbs. ( 377.8 kg)
Section 6
Specifications
720
230VAC, 1 PH, 50/60 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
15 Amp, 230VAC, Dedicated Circuit,
15 Amp Time Delay Breaker
7010708