Section 6
Specifications
6-2
7010708
Model
715
Temperature Range
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
Exterior Dimensions
72.0"W x 40.5" H x 28.8" F-B
(182.9cm x 102.9cm x 73.2cm)
Add 7.9" (20.1cm) to F-B for wall spacer/handle.
Interior Dimensions
42.5"W x 28.0"H x 18.5" F-B
(108.0cm x 71.1cm x 47.6cm)
Capacity
12.7 cu. ft. (360 liters)
Refrigeration
Cascade system, (2) hermetically-sealed compressors
Insulation
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;
1.0" (12.5cm) sub-lids
Electrical
120VAC, 1 PH, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
Breaker Requirements 20 Amp, 120VAC, Dedicated Circuit,
20 Amp Time Delay Breaker
Shipping Weight
716 lbs. (324.8 kg)
Model
718
Temperature Range
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
Exterior Dimensions
87.6.0"W x 40.5" H x 28.8" F-B
(222.5cm x 102.9cm x 73.2cm)
Add 7.9" (20.1cm) to F-B for wall spacer/handle.
Interior Dimensions
58.8"W x 28.0"H x 18.5" F-B
(149.4cm x 71.1cm x 47.0cm)
Capacity
17.0 cu. ft. (481.4 liters)
Refrigeration
Cascade system, (2) hermetically-sealed compressors
Insulation
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;
1.0" (12.5cm) sub-lids
Electrical
120VAC, 1 PH, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
Breaker Requirements 20 Amp, 120VAC, Dedicated Circuit,
20 Amp Time Delay Breaker
Shipping Weight
821 lbs. ( 372.4 kg)
712
230VAC, 1 PH, 50/60 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
15 Amp, 230VAC, Dedicated Circuit,
15 Amp Time Delay Breaker
719
230VAC, 1 PH, 50/60 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
15 Amp, 230VAC, Dedicated Circuit,
15 Amp Time Delay Breaker