Thermo Scientific Ultima II Installation And Operation Manual

Thermo scientific revco cryogenic storage freezers

Advertisement

Installation and Operation Manual
®
Thermo Scientific Revco
Ultima II
Cryogenic Storage Freezers

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific Ultima II

  • Page 1 Installation and Operation Manual ® Thermo Scientific Revco Ultima II Cryogenic Storage Freezers...
  • Page 2 © 2007 Thermo Fisher Scientific. All rights reserved. “Suva ® ” is a registered trademark of DuPont. ® “Mylar ” is a registered trademark of DuPont. “Armaflex ® ” is a registered trademark of Armacell International GmBH. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. 2007 Thermo Fisher Scientific.
  • Page 3: Table Of Contents

    ................2 6 Operation (Revco Ultima II Freezers) .
  • Page 4 Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation 8 Chart Recorders ..............8 8.1 Set Up and Operation...
  • Page 5: About Cryogenic Models

    Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation 1 About Cryogenic Models 3.1 Initial Loading After reading and completing the Safety Considerations, The freezers described in this manual are cryogenic models. Pre-Installation, Installation, and Operation sections of this These models are designed for optimum performance at -140ºC manual, turn the Key Switch to the POWER ON position and or -150ºC, and it is strongly recommended that you use these...
  • Page 6: Installation

    Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation 5 Installation Do not exceed the electrical and temperature ratings printed on 30A-250V the dataplate located on the lower left side of the unit. NEMA 6-30P CAUTION! Improper operation of the equipment could result in dangerous conditions.
  • Page 7: Operation (Revco Ultima Ii Freezers)

    11. Control Setpoint. Press this pad to set the cabinet temperature Figure 2. Revco Ultima II Series Control Panel with (refer to Section 6.2.2 on page 4). Optional Backup System 12. Cabinet Temperature Indicator. This indicator lights when the...
  • Page 8: Start Up

    Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation If a power failure lasting over 30 seconds occurs, the Note: 6.2 Start Up POWER ON light goes out and the POWER FAILURE Refer to Section 6.1 and Figure 2 on page 3 as you complete the light and digital temperature display flash following procedures.
  • Page 9: Backup System (Optional)

    Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation You can press the Audio Silence pad to silence the alarm but the 6.3.1 LN Precautions alarm will ring back in five to seven minutes. You can silence the The following are precautions for using liquid LN...
  • Page 10: Installation

    Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation 6.3.2 Installation Field installed systems are supplied with complete installation Inlet Conn. and operating instructions. If your system is factory installed, the Copper Tubing Factory Supplied freezer is shipped with a coiled length of tubing to connect the 0.625"...
  • Page 11: Maintenance And Troubleshooting

    Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation 6. If there is freezer odor, wash the interior with a solution of 7 Maintenance and Troubleshooting baking soda and warm water. Clean the exterior with any common household cleaning wax.
  • Page 12: Chart Recorders

    Thermo Scientific Revco Cryogenic Storage Freezers Installation and Operation 8 Chart Recorders 8.2 Power Supply The recorder normally uses AC power when the system is Panel-mounted six-inch seven-day recorders are available as operating. If AC power fails, the LED indicator flashes to alert options for all freezer models.
  • Page 13 Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito www.thermo.com/ Conformité...
  • Page 14 Important For your future reference and when contacting the factory, please have the following information readily available: Model Number: Serial Number: Date Purchased: The above information can be found on the dataplate attached to the equipment. If available, please provide the date purchased, the source of purchase (manufacturer or specific agent/rep organization), and purchase order number.
  • Page 15 Thermo Fisher Scientific Inc. 275 Aiken Road Asheville, NC 28804 United States www.thermofisher.com 39125H51 Rev. D...