Summary of Contents for Thermo Scientific BIOShield 720
Page 1
Thermo Scientific BIOShield™ 720 Instruction Manual 50121034-d • 07 / 2020 Visit us online to register your warranty: thermofisher.com/labwarranty...
Page 2
WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol:...
• Do not use rotors which show any signs of corrosion and / or cracks. • Use only with rotors which have been loaded properly. • Never overload the rotor. • Never run the rotor without lid. Thermo Scientific 75003621 BIOShield...
Page 8
ATTENTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur. This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. 75003621 BIOShield Thermo Scientific...
WARNING If the rotor cannot be properly locked in place after several attempts, then the ™ AutoLock is defective and you are not permitted to operate the rotor. Check for any damage to the rotor, damaged rotors must not be used. Keep the hub area clear of objects. 75003621 BIOShield Thermo Scientific...
2. Grab the rotor handle with both hands and press against the green AutoLock button. At the same time, pull the rotor directly upwards with both hands and remove it from the centrifuge spindle. Make sure not to jam the rotor while doing this. Thermo Scientific 75003621 BIOShield...
Always be sure to use buckets of the same weight class, which is marked on the buckets themselves. Proper Loading To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times. 75003621 BIOShield Thermo Scientific...
There are two options available for centrifuging samples whose weight, including adapter, exceeds the maximum permissible load: • Reduce the fill level. • Reduce the speed. Use the table or the formula: Actual Load (g/bucket) 6300 6170 6048 5933 5824 5721 5623 5530 5442 5357 5277 5200 5126 5055 Thermo Scientific 75003621 BIOShield...
WARNING Replace the rotor when the specified number of cycles is reached. Due to the mechanical load, a rotor can break and thus damage the centrifuge. Service life examples Usage profile Maximum service life at 30,000 cycles frequent use 7 years 20 runs / day 220 days / year 75003621 BIOShield Thermo Scientific...
Check as needed to make sure your rotor is aerosol-tight. The careful inspection of the seals and seal surfaces for signs of wear and damage such as cracks, scratches and embrittlement are extremely important. Aerosol-tight applications are not possible if the lids are open. 75003621 BIOShield Thermo Scientific...
Page 29
Damaged or clouded rotor covers are to be replaced immediately! CAUTION This quick test is not suited for validating the aerosol tightness of your ™ BIOShield 720. Check the seals and sealing surfaces of the lid thoroughly. Thermo Scientific 75003621 BIOShield...
• Store the aluminium parts at room temperature or in a cold-storage room with the cavities pointing down. CAUTION Before using any cleaning or decontamination methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. 75003621 BIOShield Thermo Scientific...
CAUTION Before using any cleaning or decontamination methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. Observe the safety precautions and handling instructions for the cleaning agents used. Thermo Scientific 75003621 BIOShield...
For general radioactive decontamination use a solution of equal parts of 70 % ethanol, 10 % SDS and water. Disinfect the rotor and accessories as follows: 1. Open the centrifuge. 2. Turn off the centrifuge. 75003621 BIOShield Thermo Scientific...
Place the rotor on a flat surface. Note No chemical additives are permitted in the steam. CAUTION Never exceed the permitted temperature and duration when autoclaving. If the rotor shows signs of corrosion or wear, it must be replaced. Thermo Scientific 75003621 BIOShield...
• the suitability of the set-up site; • the lid lock and the safety system; • the rotor; • the fixation of the rotor and the drive shaft. Thermo Fisher Scientific offers inspection and service contracts for this work. 75003621 BIOShield Thermo Scientific...
S U U U U U U U - M S M - Boric Acid U S S M S S U S S S S S S S S S U S S S S S S S S S S Thermo Scientific 75003621 BIOShield...
Page 40
S S S U S U S S S U ® M S S - S S S - S S S S - S S - S S S S S S S M S S S Ficoll-Hypaque 75003621 BIOShield Thermo Scientific...
Page 41
S S S U S S M S S S S S U S U M S S S S S S S M S M U U U U U M S S U S U U S U U U U U M U U U S S M U S U Methylene Chloride Thermo Scientific 75003621 BIOShield...
Page 42
S S S S S S S S S S S S S S S S S S S S S U - S M S S S M S U S S S U S S S S U S U M S M S Picric Acid 75003621 BIOShield Thermo Scientific...
Page 43
S S S S S S S S S S S S S S S S S S S S S S S M S S M S S M S S S S S S S S S M S S S S S S S S S S Citric Acid (10 %) Polyethyleneterephthalate Thermo Scientific 75003621 BIOShield...
Page 44
Performance unknown; suggest testing, using sample to avoid loss of valuable material. Chemical resistance data is included only as a guide to product use. Because, no organized chemical resistance data exists for materials under the stress of centrifugation, when in doubt we recommend pretesting sample lots. 75003621 BIOShield Thermo Scientific...