10. Open all flaps same way. Remove any and all visible soiling, using approved
disinfectant.
11. Remove all grommets:
12. Clean the grommets with a disinfecting solution.
13. Remove the mattress from inside the incubator:
14. Wipe mattress down with approved cleaner and check for holes or punctures in
material.
15. Remove the mattress tray very CAREFULLY so as not to damage the sensor unit
below: