Guidance and manufacturer's declaration – Electromagnetic Emissions
T CaRe Compact is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment.
Emission test
RF Emissions
Cispr 11
RF Emissions
Cispr 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user should assure that it is used in such an environment.
Immunity test
Electrostatic
±8kV contact
discharge
±2-4-8-15 kV air
(ESD)
4 kV indirect contact
EN 61000-4-2
Burst/Fast
±2kV power supply
Transient
lines
EN 61000-4-4
Surge
±0,5-1kV differential
EN 61000-4-5
mode
±0,5-1-2kV common
mode
Voltage
dips,
Voltage reduction 100%
short
(10ms)
interruptions and
Voltage
voltage variations
(500 ms)
on power supply
Voltage reduction 100%
input lines
(5000 ms)
EN 61000-4-11
Power frequency
30 A/m
magnetic field
EN 61000-4-8
NOTE:U
is the a.c. mains voltage prior to application of the test level
T
Easytech s.r.l. – User Manual T CaRe Compact
Conformity
T Care must emit electromagnetic energy in order to perform its
Group 2
intended function. Nearby electronic equipment may be affected.
T-Care is suitable for use in all establishments, excluded domestic
establishments and those directly connected to the public low-
Class A
voltage power supply network that supplies buildings used for
domestic purposes.
T CaRe Compact is suitable for use in all establishments, including
Class A
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
T CaRe Compact is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
Complies
low-voltage power supply network that supplies buildings used for
domestic purposes.
Tabella 2-1 – EMC, Emissions
Test level
EN 60601-1-2
±8kV contact
±2-4-8-15 kV air
4 kV indirect contact
±2kV power supply
lines
±0,5-1kV differential
mode
±0,5-1-2kV common
mode
Voltage
(10ms)
reduction
30%
Voltage reduction 30% (500
ms)
Voltage
(5000 ms)
30 A/m
Tabella 2-2 – EMC, Immunity
Electromagnetic Environment - guidance
Compliance Level
reduction
100%
reduction
100%
rev. D
[tab.1/60601-1-2]
[tab.2/60601-1-2]
Electromagnetic Environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user requires
continued operation during power
mains interruptions, it is
recommended that the device be
powered from an uninterruptible
power supply or a battery.
Magnetic power frequency field
should
be
that
of
commercial
or
environment.
Pag. 9 di 36
a typical
hospital