Guidance and manufacturer's declaration – Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user should assure that it is used in such an electromagnetic environment.
Immunity test
Test level
EN 60601-1-2
3 Veff from
150kHz to
80MHz
6 Veff from
6,765MHz to
RF conducted
6,795MHz, from
EN 61000-4-6
13,553MHz to
13,567MHz, from
26,957MHz to
27,283MHz, from
40,66MHz to
40,70MHz
3 Veff from
150kHz to
80MHz
6 Veff from
6,765MHz to
6,795MHz, from
13,553MHz to
13,567MHz, from
26,957MHz to
27,283MHz, from
40,66MHz to
RF radiated
40,70MHz
EN 61000-4-3
3 Veff/m2 from
80MHz to
2700MHz
27 Veff/m2 from
380MHz to
390MHz
28 Veff/m2 from
Easytech s.r.l. – User Manual T CaRe Compact
Compliance
Level
Portable and mobile RF communications equipment
3 Veff from
should be used no closer to any part of
150kHz to
including cables, than the recommended separation
distance calculated from the equation applicable to the
80MHz
frequency of the transmitter.
Recommended separation distance
d = 1,2 .√P from 150kHz to 80MHz
6 Veff from
d = 1,2 .√P from 80 MHz to 800 MHz
d = 2,3 .√P from 800 MHz to 2,5 GHz
6,765MHz to
where P is the maximum output power rating of the
6,795MHz, from
transmitter in watts (W) according to the transmitter
13,553MHz to
manufacturer and d is the recommended separation
13,567MHz,
distance in metres (m)
from 26,957MHz
Field strengths from fixed RF transmitters, as determined
to 27,283MHz,
by an electromagnetic site survey
from 40,66MHz
the compliance level in each frequency range
to 40,70MHz
Interference may occur in the vicinity of equipment
marked with the following symbol:
Portable and mobile RF communications equipment
3 Veff from
should be used no closer to any part of
150kHz to
including cables, than the
recommended separation distance calculated from the
80MHz
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1,2 .√P from 150kHz to 80MHz
6 Veff from
d = 1,2 .√P from 80 MHz to 800 MHz
d = 2,3 .√P from 800 MHz to 2,5 GHz
6,765MHz to
where P is the maximum output power rating of the
6,795MHz, from
transmitter in watts (W) according to the transmitter
13,553MHz to
manufacturer and d is the
13,567MHz,
recommended separation distance in metres (m)
from 26,957MHz
Field strengths from fixed RF transmitters, as determined
to 27,283MHz,
by an electromagnetic site survey
from 40,66MHz
the compliance level in each frequency range
to 40,70MHz
Interference may occur in the vicinity of equipment
marked with the following symbol:
3 Veff/m2 from
80MHz to
2700MHz
27 Veff/m2 from
380MHz to
390MHz
28 Veff/m2 from
rev. D
[tab.4/60601-1-2]
Electromagnetic environment - guide
the device
,
a
, should be less than
b
.
the device
,
a
, should be less than
b
.
Pag. 10 di 36