easytech T CaRe Compact User Manual page 23

Table of Contents

Advertisement

The operator can work by setting the treatment length (
energy
supplied (
mode).
If the operator has set the duration of treatment (time mode) operation, the appliance
displays the value of the energy supplied (in the center of the screen, see Fig. 4.1, this value,
then, will be increased during treatment).
If it is set the value of energy to be supplied, the machine calculates the treatment duration.
(see. Figure 4.2)
Figure 4-2 – Manual setting of operating parameters –Energy mode
After defining the parameters, follow these steps:
1. place the electrodes in the area to be treated
2. give the start command
3. Set the power to be delivered
start
The
command can also be given through the button on the
When the treatment begins (see Figure 4-3), the timer starts counting backwards, indicating
the time residual; it is also activated an audible signal (
 In the
time mode
 In the
energy mode
On the screen, in the left bar, are shown the power level (Figure 4-3) and the current
(shaded area in Figure 4-3); it is also shown a value that represents the electrical resistance
of the tissue.
The operator can use the Touch screen to
knob.
1
In reality the value of the voltage supplied by the device is programmed; for simplicity, it is called power.
Easytech s.r.l. – User Manual T CaRe Compact
(in % of the
1
the treatment will stop when the timer reaches 0.
treatment will stop on reaching the set value.
stop
time
mode) or the energy to be
maximum).
knob.
beep
).
the treatment or press the button on the
rev. D