5.5 Operating in Capacitive Mode

By using the electrode for the capacitive operational modality, you can treat superficial
parts, muscles and superficial vessels (thus skin and muscular diseases).
In order to distribute the heat uniformly, the mobile electrode should be moved with
appropriate movements, keeping it parallel to the fixed electrode.
Important: the entire electrode surface should rest flat on the patient's skin to avoid a
concentration of current on a small area, which could have harmful effects.
Treatment is started with a low power level, which is then increased until the patient
doesn't feel a warm sensation; the value of power must be reduced as soon as the patient
perceives excessive heat.

5.6 Checks to be performed during the treatment

During treatment the operator should check to make sure that the patient does not feel
excessively hot in correspondence with the reference electrode. If this condition occurs, the
treatment should be discontinued and the electrode should be smeared with conductive
cream and possibly repositioned.