7. MAINTENANCE

7.1 Cleaning the exterior

 Switch off the equipment by pressing the button in the back.
Disconnect the 220V power supply plug from the wall socket of the room. On this
occasion, also inspect the cable (see section 2.3).
 The handpiece and the base of the unit can be cleaned with a cloth moistened with
water or not aggressive detergent / disinfectant, but there must not be free liquid
anywhere during cleaning.
 Avoid splashing of liquid especially under the front, against which the equipment does
not have a specific protection.
 Do not use solvents or other solutions containing abrasives.
 Do not use heat / pressure sterilization systems.
The electrodes must be cleaned after use. They can be cleaned with a solution
detergent / disinfectant; for the disinfection of metal electrodes a product of hospital
use for stainless steel can be used. Before use, make sure the electrodes are
completely dry.

7.2 Periodic inspection and life cycle of the unit

The most important aspect to be kept under control is the electrical safety, because over
time increases the likelihood that some event may occur which may adversely affect, for
example, an electrical insulation.
As a basic rule it is recommended an annual inspection by a specialized laboratory which
can also make the performance at home. The interval of a year is the most commonly used
for devices of this class, however, periodic checks can be performed in accordance with the
choices of quality policies by each Center. The reference standard for the electrical safety
testing is EN 62353.
As already suggested in the section of the requirements for security, we emphasize the
usefulness of carrying out simple visual / tactile inspections whenever the unit is used to
make sure that in any case the machine falls in safe operating condition and has not
suffered any visible injuries.
In the aim of preventive maintenance, we suggest a general check by the manufacturer after
5 years (or in any case every five years at most), in particular to check whether there have
been events "traumatic" that, while allowing the overcoming of electrical tests, may however
have reached a critical point for a sudden collapse.
We consider that
after 10
the construction, the materials used and the maintenance schedule lead to foresee a safety
behavior over a far longer period. This is obviously a purely indicative matter since it strongly
depends on the way of application of the device.
Easytech
ensures the maintenance of the unit up to 5 years after its exit from the market
(according to existing law).
Easytech s.r.l. – User Manual T CaRe Compact
years of work the equipment has terminated his life cycle, even if
rev. D