T Care Compact Technical Characteristics - easytech T CaRe Compact User Manual

Table of Contents

Advertisement

3.4 T CaRe Compact Technical Characteristics

Classification according to Directive 93/42/EEC
Classification according to EN 60601-1 standard
Control unit microprocessor based
Touch screen monochromatic display, high brightness
Working frequencies
Output power RES (at 82 ohm)
Output power CAP (at 1000 ohm)
Maximum output voltage on the resistive electrode (open circuit)
Maximum output voltage on the
circuit)
Power supply (voltage and frequency)
Maximum power consumption
Electric protection
Dimensions WxDxH
Weight
Environment temperature
Relative humidity (non-condensing)
Atmospheric Pressure
Easytech s.r.l. – User Manual T CaRe Compact
capacitive
electrode (open
rev. D
Medical Device IIb
Class I device, with applied part:
BF
130x72mm
500 kHz ±5%
200 W
250 VA
340 Vpp
500 Vpp
100-230 Vca, 50 / 60 Hz
350 W
2 Fuses of 4 A
225x300x185 mm
4,4 Kg
+ 10 ÷ + 40°C
0 ÷ 75%
700 ÷ 1060 mbar