Peak Scientific Precision 500 User Manual

Peak Scientific Precision 500 User Manual

230v hydrogen gas generator
Hide thumbs Also See for Precision 500:

Advertisement

Quick Links

Precision 500 230v
Hydrogen Generator
User Manual
2012 © Peak Scientific – Rev 2 – 07/11/12

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Precision 500 and is the answer not in the manual?

Questions and answers

Summary of Contents for Peak Scientific Precision 500

  • Page 1 Precision 500 230v Hydrogen Generator User Manual 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 2: Table Of Contents

    Supply Line Tubing Switching the unit off Cleaning Replacement of Deioniser Resin Column Refilling the Deioniser Resin Column Health and Safety Data- Deioniser Resin IMPORT ANT DOCUMENTS Service Requirements Trouble Shooting Page 2 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 3: Change History

    Please review each of the following sections carefully. Thank you for selecting Peak Scientific to meet your Gas Generation needs, and should you require any further assistance or support please do not hesitate to contact Peak Scientific or Peak Partner from which you purchased your Generator.
  • Page 4: Introduction

    Precision 500 User Manual Introduction The Precision 500 has been developed to cater for the requirements of GCMS instruments. This model provides a source of High Purity Hydrogen gas. Other features include: • Touch screen interface • >99.9999% purity •...
  • Page 5: Warranties And Liabilities

    (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party. Page 5 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 6: Safety Notices

    Safety Notice to Users These instructions must be read thoroughly and understood before installation and operation of your Peak Precision 500 Hydrogen Generator. Use of the Generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment.
  • Page 7 ATTENTION - IMPORTANT NOTICE TO ALL USERS These instructions must be read thoroughly, section by section and understood before installation and operation of your Peak Scientific Hydrogen Generator. Please follow the appropriate safety standards for handling hydrogen gas and other gases as determined by the laws and regulations of your country.
  • Page 8: Declaration Of Conformity

    York EMC Services (2007) Ltd Donibristle Industrial Park Dunfermline Fife KY11 9HZ Signed By: Name: Chris Pugh Position: Engineering Director Done at: Peak Scientific Instruments Ltd, Inchinnan, Scotland. Date: August 2012 Page 8 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 9: Environmental Declaration

    Where a specific certificate of compliance is required, this can be requested, on a product serial number basis, directly from Peak Scientific Instruments Ltd, by contacting us through our website on www.peakscientific.com...
  • Page 10: Technical Specification

    Shipping crate dimensions in cm W x D x H 55 x 80 x 68 (21.6” x 31.4” x 26.7”) Shipping weight 66 kg (145 lb) Water purity requirement <1.0µ Siemens/cm OR >1 Mohm-cm Page 10 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 11: Principle Of Operation

    NASA. To produce the highest purity of hydrogen that an analytical laboratory requires, Peak Scientific employs the services of a platinum catalyst in the electrode area.
  • Page 12: General Safety

    Precision 500 User Manual General Safety The construction of the Precision 500 generator conforms to CEI / OSHA regulations as they do not store any appreciable amount of hydrogen internally. The generator only produces gas when the application places a demand on the unit.
  • Page 13: Generator Safety Features

    Precision 500 User Manual Generator Safety Features The Precision 500 hydrogen generator has the following safety features to ensure safe and reliable operation: To avoid any build up of hydrogen from leaks inside the unit; the generator features forced air ventilation. This ensures that a hydrogen / oxygen mixture cannot form inside the generator.
  • Page 14: Unpacking & Contents

    Any damage should be reported immediately to the carrier and Peak Scientific or the distributor from whom the unit was purchased. If possible, unpack the generator and allow to stand at room temperature for a few hours.
  • Page 15: Installation

    Minimum Operating Ambient Temperature: 10 °C (50 °F) Maximum Operating Ambient Temperature: 25 °C (77 °F) Note: It is recommended that the unit is left to run for 12 hours before utilisation. Page 15 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 16: Unpacking

    Precision 500 User Manual Unpacking Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage. Check ‘SHOCKWATCH’ label for signs of rough handling prior to un-packing –...
  • Page 17: General Dimensions

    Precision 500 User Manual General Dimensions The Generator must always be placed on a flat, level surface. Failure to do so will affect the performance of the Generator. Page 17 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 18: Unit Controls

    Master Connection (Daisy Chain) Exhaust O Vent H Vent Slave Connection (Daisy Chain) Detector Connection Output Fuse Power Power Water Supply Drain Inlet Switch Fill Figure 2: Rear Connections Page 18 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 19: Tubing

    In addition, plastic tubing can give off organic impurities which can cause ghost peaks and baseline instability. Furthermore Peak Scientific recommend that Swagelok® fittings are used, and that the fittings are free from Teflon® tape.
  • Page 20: Automatic Filling

    Using the length of Tygon® tubing and the hose barb fitting from the fittings kit, attach the barb to the tubing. Then connect one end to the water supply and the other to the Fill port at the rear of the Precision 500. Page 20...
  • Page 21: Electrical Connection

    Green LIVE (L):- Brown Black Neutral (N):- Blue White Our electrical requirements are 230VAC nominal +/- 10%. This means that the Generator can accommodate transients between 207VAC and 253VAC. Page 21 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 22: Start-Up Sequence

    Peak service provider. Liquid Leak Detectors Liquid leak detectors should not be used on the Precision 500. These solutions are easily drawn into the high purity hydrogen gas system and can cause irreparable damage to a GC column. If a liquid solution has to be used, then it should be IPA based, or a portable hydrogen detector should be used.
  • Page 23: H 2 Detector Option

    Precision 500 will alarm if dangerous levels of hydrogen are detected, the Precision 500 will then shut down. Multiple detectors can be connected to one Precision 500, up to 4 in number. If this is desired, connect the first detectors ‘Out’ port to the second detectors ‘In’...
  • Page 24: Gas Traps

    Peak generator and the instrument to provide back-up contaminant protection for your instrument. The hydrogen gas produced by the Precision 500 will be sufficiently pure for these contaminant traps to last a long time, but should a leak ever occur on the gas line, the traps will ensure that any harmful contaminants are caught, therewith preventing potential damage to the instrument.
  • Page 25: Expanding Capacity

    Rear Connections In order to use this feature the units must be connected together via the ‘Master’ port on the first Precision 500, and the ‘Slave’ port at the rear of the next Precision 500 unit. The first Precision 500 unit is automatically defined as the Master, and before...
  • Page 26: Supply Line Tubing

    Precision 500 User Manual Supply Line Tubing Peak scientific recommends using only clean grade copper or stainless steel for the supply tubing. The ring main and series lines should be ¼” OD with the T- pieces 1/8" on the incoming connections from the generators and 1/4" on the main line connections.
  • Page 27: Switching The Unit Off

    Generator. Under no circumstances should any solvents or abrasive cleaning solutions be used as these can contain fumes that could be harmful to the Generator. Page 27 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 28: Replacement Of Deioniser Resin Column

    5. Remove the used column by pulling it towards you. 6. Fit a re-filled Deioniser Column into the generator and reconnect the tubing, 7. Replace the cover and screws. 8. Switch the unit on. Page 28 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 29: Refilling The Deioniser Resin Column

    O ring and screw the lid back on the column, until the O ring is in contact with the cartridge. Ensure the lid is not over tightened, as this could damage the thread. Page 29 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 30: Health And Safety Data- Deioniser Resin

    Fire & Explosion Hazards Toxic combustion products may include sulphur dioxide and sulphur trioxide Disposal This product can be disposed as ordinary waste if local regulations permit Page 30 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 31 No toxicity data available on this product Occupation exposure limits STEL/TLV = None established. Incompatible Materials (to avoid hazardous reactions) Avoid contact with concentrated nitric acid and any other strong oxidising agents at all times. Page 31 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 32: Important Documents

    MODEL TYPE: INSTALLATION DATE (DD/MM/YYYY): TELEPHONE: Important Please Note: You have 1 month to register your Peak Scientific product from the date of shipment. If you wish to defer installation of your generator you must notify Peak Scientific within month shipment date.
  • Page 33: Service Requirements

    2 x Deioniser Resin Refill Pack 2 x Column Sponge Service Plans Peak Scientific offer two service plans. The Complete Service Plan, specifically designed for Generators operated in critical environments, also includes full breakdown cover, guaranteed response times and Generator upgrades if available.
  • Page 34: Trouble Shooting

    • Check external pipework for leaks. Capacity Excess Alarm • Contact your service provider. Water Level Fault • Contact your service provider. Hydrogen Detected • Check for external leaks. Page 34 2012 © Peak Scientific – Rev 2 – 07/11/12...
  • Page 35 New Taipei City São Paulo SP Brasil 107-0061 Singapore, 118223 Tel: +886 2 8226 2383 Tel: +1 866 647 1649 Tel: +81 3-6864-0468 Tel: +65-6777 1966 +886 2 8226 9499 Fax: +65-6777 1978 2012 © Peak Scientific – Rev 2 – 07/11/12...