Installation Guide – Precision 500 Generator 110V
x 1m
" Tygon Tube
2.5mm Hex Key
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

FITTINGS KIT

Mains Cable – US
x 1
Resin Column Re-fill
CD containing User Manual
x 1
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E:
info@peakscientific.com
x 1
x 1
Barbed Fitting
Part No. IG-Precision
Rev.
01
Page
Page 1 of 8

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Precision 500 and is the answer not in the manual?

Questions and answers

Summary of Contents for Peak Scientific Precision 500

  • Page 1 Installation Guide – Precision 500 Generator 110V FITTINGS KIT Mains Cable – US x 1m ⅜ ” Tygon Tube CD containing User Manual 2.5mm Hex Key Resin Column Re-fill Barbed Fitting Fountain Crescent T: +44 (0)141 812 8100 Part No. IG-Precision...
  • Page 2 INSTALLATION Ensure the ¼” and ⅜” plugs have been removed Unpack the generator from the shipping crate and position on a flat from the O and H vents, located at the rear of the surface, in desired area. generator. Connect the ⅜” tubing, contained in the fittings Connect the other end of the tubing to the kit, to the H Delivery port at the rear of the...
  • Page 3 Connect an external supply of deionised water to the Fill connection at the rear of the Select the mains cable from the fittings kit and unit, using the supplied tubing. Water plug the IEC 60320/C13 socket into the mains loading takes place automatically by means input at the rear of the generator.
  • Page 4 The touch HMI screen will illuminate and The system will now run diagnostics and a the start screen will be displayed. leak detection test. * NOTE * If diagnostics are unsuccessful, an alarm will sound and a red failure screen will be displayed, complete with a mute button.
  • Page 5 Once the desired unit of measure and Pressing START will start the supply to the pressure have been selected, pressing application. The unit may take a few minutes BACK will return the user to the home to adjust to the desired pressure. screen.
  • Page 6 CD supplied in the fittings kit. Please keep this for future reference. It is very important to register your generator with PEAK SCIENTIFIC. This will initiate your warranty entitlement. Please use the form on the next page to register your generator.
  • Page 7 You have 1 month to register your Peak Scientific product from the date of shipment. If you wish to defer installation of your generator you must notify Peak Scientific within 1 month of the shipment date. This can be done by emailing warranty@peakscientific.com...
  • Page 8 If multiple Precision 500 generators are being installed they must be connected as shown in the image below. The units must be connected together via the ‘Master’ port on the first Precision 500, and the ‘Slave’ port at the rear of the next Precision 500 unit.