Calibration Curve Set A For General Bulk Solids Applications - Endress+Hauser Solitrend MMP60 Operating Instructions Manual

Material moisture measurement
Table of Contents

Advertisement

Solitrend MMP60
• CK mode (Cyclic with Boost Filter)
For complex applications in mixers and dryers
• CC mode (Cyclic Cumulated)
With automatic totalization of moisture quantity measurements in one batch process if no
PLC controller is used
• CH mode (Cyclic Hold)
Standard operating mode for applications in the construction industry. Similar to the CC
mode, but with filtering and without totalization. The CH mode is ideal for very short batch
times down to 2 s if the sensor has been installed under the silo discharge hatch. The CH
mode performs filtering automatically. This allows drip water that forms in the silo to be
filtered out of the measured value, for example.
8.3

Calibration curve set A for general bulk solids applications

The devices are supplied with a suitable calibration. A maximum of 15 different calibrations
can be saved in the device and can be activated and adjusted via the remote display. To pre-
test the compatibility of a calibration curve, the user can select individual calibration curves
(Cal.1 to Cal.15) in the Material cal. menu item, test the curve with the material to be
measured and activate it. The desired calibration curve - which may have been modified - is
active once the operating voltage is switched on.
Endress+Hauser
Commissioning
15