Electrical connection
Is the device undamaged (visual inspection)?
If provided: are the measuring point number and labeling correct?
Are the connections established correctly and protected against mechanical influences?
If used: is the device securely positioned in the mounting flange / mounting frame (visual
inspection)?
Is the device securely mounted and is the measuring cell surface flush on the material side
(visual inspection)?
Is sufficient material coverage / material flow over the measurement surface ensured?
6
Electrical connection
6.1
Connecting requirements
6.1.1
Cable specification
Connecting cables with a pre-assembled 10-pin socket on the device side are available in
different standard lengths:
• 4 m (13 ft)
• 10 m (32 ft)
• 25 m (82 ft)
Shielded cable UNITRONIC PUR CP, twisted pairs 6 × 2 × 0.25 mm
oils and chemicals.
On request: Shielded cable UNITRONIC ROBUST CP 10 × 0.25 mm
oils and chemicals.
6.2
Connecting the device
6.2.1
Terminal assignment
The device is supplied as standard with a 10-pin plug, series 26482, with IP67 protection.
In the case of the electronics module located remotely by means of the HF cable, the
electronics housing is fitted on both sides with plugs with IP67 protection.
10
Solitrend MMP60
2
, PUR sheath resistant to
2
, PUR sheath resistant to