4.3 Software initiation

Turn on the machine following the instructions pasted on the wall and inserted in the log book.
Wait for complete initialization of microscope and Windows after logging in with your TLL
username and password.
Double click on:
(A) The LAS X configuration screen will appear:
Configuration: Machine.xlhw must be selected.
Microscope: Check that DMI6000 is selected.
Click on OK and wait for complete initialization. Do not move anything on the microscope
during the initialization process.

4.4 Turning on the lasers

1. Under the Visible panel, click on the
2. Click on the UV (405nm), Visible (Argon) and SuperContVisible2(WLL) bar to observe
405nm, Argon and WLL respectively in the dropdown menu.
3. White light laser (WLL) has excitation wavelengths from 470nm-670nm allowing you to
select specific excitation light source. Argon lasers have much higher power than the WLL
and has excitation wavelengths – 458nm, 476nm, 488nm, 496nm and 514nm.
4. Turn on the laser that you need by clicking on the sliders. Make sure that WLL and
SuperContVisible2 are at 70% before carrying on with the next steps.
1

4.5 Using Dye Assistant tab

• Select the objective lens that you would like to image with your sample.
button to open the laser panel.