Installation Procedure - Agilent Technologies PSG Series Installation Notes

Signal generators add option 1e1 attenuator upgrade kit
Hide thumbs Also See for PSG Series:
Table of Contents

Advertisement

Installation Procedure

Remove the RF Cable from the Coupler Detector to the RF Output Connector
1. Position the signal generator with the A31 Motherboard facing you and the RF deck on top.
2. Using a 5/16" open wrench, remove the semi-rigid RF cable connecting the A23 Coupler Detector to the
RF output connector.
Installing the AT1 115 dB Attenuator to the RF deck
Refer to
Figure
3.
1. Position the attenuator (AT1) so it can be attached to the RF deck.
2. Using the T-10 driver, insert the two screws (1) through the side deck into the metal bracket that is
attached to the attenuator. Torque these screws to 9 in-lbs.
3. Using the 5/16" open ended wrench, attach the semi rigid cable W24 between the A23 Coupler Detector
output connector and the attenuator input connector; torque to 8 in-lbs.
• 20 GHz models : use E8251-20026
• 40 GHz models: use E8251-20025
4. Using the 5/16" open ended wrench, attach the semi rigid cable W25 between the A23 Coupler Detector
output connector and the 3.5mm RF output connector; torque to 8 in-lbs.
• 20 GHz models : use E8251-20027
• 40 GHz models: use E8251-20028