Related Adjustments; In Case Of DiCulty - Agilent Technologies 8360 B Series Service Manual

Swept signal/cw generator
Hide thumbs Also See for 8360 B Series:
Table of Contents

Advertisement

3. On the oscilloscope, set:
Channel 1:
Display
Volts/Division
Input Coupling
Input Impedance 50
Timebase:
Time/Division
Delay Reference center
weep
Trigger:
Trigger Mode
Trig rc
Trigger Level
lope
Display:
Display Mode
Averaging
Display Time
4. Adjust the oscilloscope o set to fully display the signal.
5. Use the oscilloscope to measure the width of the pulse. Record the value measured on the
test record.
6. On the synthesizer, change the pulse width to 50 ns. Press:
Internal Pulse Width
7. Use the oscilloscope to measure the width of the pulse. Record the value measured on the
test record.

Related Adjustments

None
In Case of Difficulty
Refer to the
Agilent Technologies 8360 B-Series Swept Signal Generator/8360 L-Series
Swept CW Generator Troubleshooting Gui e
2-106 Performance Tests
Internal Pulse Accuracy
On
500 mV
dc
10 ns
auto
Edge
Chan 1
1.2V
Pos
Repetitive
Off
0.2 s
50
ns
5
5
.
Agilent 8360