6. Note the nal center frequency setting required to center the video signal.
7. Record the dierence between the initial center frequency setting and the value noted in
step 6 in able 2-2 as frequency error.
8. Repeat steps 5 through 7 for the remaining instrument settings in able 2-2 and able 2-3
through able 2-6 as appropriate for each instrument model.
9. Calculate the frequency error as a percent of the sweep width as follows for each of the
results in able 2-2 and able 2-3 through able 2-6:
10. Record the worst case value on the test record.
For 83620B/22B/24B and 83623B/L only:
ynthesizer
Frequency (GHz)
tart
2
2
2
2
2
2
Agilent 8360
Figure 2-3. Video ignal on the Oscilloscope
frequency error
0
stop frequency
start frequency
Table 2-3. Additional Instrument ettings
pectrum Analyzer
Center Frequency
(GHz)
top
20
2.10125
20
6.8825
20
7.1525
20
13.25
20
13.79
20
19.775
x
100 = percent
Oscilloscope
Frequency
Trigger Events
Error
10
435
459
1001
1049
1581
wept Frequency Accuracy
Percent
Performance Tests
2-15