At the message
VERIFY BE TNOTE
wave frequency with the frequency domain signal near the left edge (0 Hz).
7. When the measurement is complete record the results on the test record and compare them
to the speci cation.
8. Repeat steps 3 through 6 for the frequencies in able 2-16.
Table 2-16. Frequency etting for Phase Noise Measurements

Related Adjustments

Sampler Assembly
In Case of Difficulty
1. Line spurs may be present in the trace which exceed the phase noise speci cation. hey
should be ignored.
2. Be sure that the signal generator frequency (detector input) is tuned close enough that a
zero beat can be found.
3. Note where the problem occurs (carrier frequency and o set frequency from the carrier)
and refer to the
Agilent Technologies 8360 B-Series Swept Signal Generator/8360 L-Series
Swept CW Generator Troubleshooting Gui e
Agilent 8360
<
, tune the signal generator for minimum sine
1 MHZ
ynthesizer
Phase Noise Measurement ystem
Frequency
Carrier Frequency
(GHz)
(GHz)
1.97
1
1.97
18.0
18.0
Disregard if beyond your synthesizer's capability.
1
Detector Input
(MHz)
310
720
.
ingle ideband Phase Noise
Performance Tests
2-51