Table of Contents

Advertisement

4. Set the default values given in able 3-4 into the corresponding ampli er/ lter
calibration constants. On the synthesizer, set:
Adjust Menu Calib Menu
SERVICE
5
Enter the number of the rst calibration constant from able 3-4 and
elect Cal
terminate the entry with
pressing the arrow keys will allow you to vary the calibration constants. he calibration
constant shown at the
Enter the default value from able 3-4 and terminate the entry with
Modify Cal
.
ENTER
5
Using the up/down arrow keys, select each of the rest of the calibration constants in
able 3-4 and set their default values by entering the value on the numeric keypad.
erminate your entry with the
3-38 Adjustments
Amplifier/Filter Adjustments
Figure 3-19. XA10J2 Location
. When using the front panel emulator software,
ENTER
5
of the \display" window is the active constant.
top
key.
ENTER
5
Agilent 8360