Agilent Technologies 8360 B Series Service Manual page 453

Swept signal/cw generator
Hide thumbs Also See for 8360 B Series:
Table of Contents

Advertisement

1. Preliminary Steps
You can easily blow the fuse for the 13.5 V supply on the preregulator
CAUTION
assembly if you do not disconnect the AC power cord. Also, the danger of
shock is increased if the power cord is connected.
1. Disconnect the AC power line cord.
2. Remove the screw from each of the four feet on the rear panel.
3. Remove the four feet.
4. o remove the side straps, remove the two screws on each side panel.
5. Remove the instrument top, bottom, and side covers.
An RF braid is in each of the side channels on the top and bottom of the
Note
instrument. When you reassemble the instrument, be sure the braids remain
in the side cover channels.
Agilent 8360
Disassembly and Replacement Procedures
7-23
Rear Panel

Advertisement

Table of Contents
loading

Table of Contents