Infrared forehead thermometer
Table of Contents

Advertisement

Acoustic:
The unit is turned on and ready for the measurement: 1 short beep. Complete the measurement: 1 long
beep (1 sec.) if the reading is less than 37.5 °C / 99.5 °F, 10 short «beep» sounds, if the reading is equal
to or greater than 37.5 °C / 99.5 °F. System error or malfunction: 3 short «bi» sounds.
Memory: 30 readings recall in the memory mode with a record of both time and date.
Backlight: The display light will be GREEN for 5 seconds, when a measurement is completed with reading less
than 37.5°C/99.5°F but equal to or higher than 34.0°C/93.2°F. The display light will be RED for 5
seconds, when a measurement is completed with a reading equal to or higher than37.5°C/99.5°F but
equal or lower 43.0°C/109.4°F .
Operating conditions: Body mode: 15-40.0 °C / 59-104.0 °F, Object mode: 5-40.0 °C / 41-104.0 °F 15-95%
relative maximum humidity
Storing conditions: -25 - +55 °C / -13 - +131 °F / 15-95% relative maximum
Automatic Switch-off: Approx. 30 sec when power on without measurement & after measurement
Battery:
2 x 1.5 V alkaline batteries; size AAA
Battery Life: Approx. 2000 measurements (using new batteries)
Dimensions: 140 x 42.2 x 50 mm
95.8 g (with batteries), 73.5 g (w/o batteries)
Weight:
IP Class:
IP22
Reference to standard: ASTM E1965; IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: 5 years or 12000 measurements
This device complies with the requirements of the Medical Device Directive 93/42/EEC. Technical alterations reserved.
According to the Medical Product User Act a biennial technical inspection is recommended for professional users. Please
observe the applicable disposal regulations.
16. www.microlife.com
Detailed user information about our thermometers and blood pressure monitors as well as services can be found at
www.microlife.com