Microlife FR1DQ1-B User Manual page 12

Infrared forehead thermometer
Table of Contents

Advertisement

If no message showed on App to confirm which patient log the data should be assigned to, please record the temperature data
showed on FR1DQ1-B into the App manually or takes a new measurement and start Transfer Data step from 1 to 6 again.
Bluetooth® Icon Indicator
The Bluetooth® icon indicator on your device, located in the middle right portion of the display, is designed to provide
information about the connection between your device and smartphone.
 The Bluetooth® function is enabled: The Bluetooth® icon indicator
 Connecting with a device or uploading the data: The Bluetooth® icon indicator
 The Bluetooth® function is automatically switched off during measurement; after measurement the device will switch
it on and upload the data to the «Microlife Connected Health+» App.
NOTES
The Bluetooth® of this thermometer is a medical device data system (MDDS) as its only function is for records and no
additional functions. It does not modify the data or modify the display of the data, and it does not by itself control the
functions or parameters of any other medical device. It is not active when the thermometer is recording data or during
measurement. The thermometer will not sound any alarm with or without Bluetooth®. The Bluetooth® is used only to
transfer data from point A to point B. The App on your smart devices cannot be used to start or stop the thermometer, nor
update the firmware of thermometer via Bluetooth®.
Wireless communication interference
This product operates in an unlicensed ISM band at 2.4 GHz. In the event this product is used near other wireless devices
such as microwave and wireless LAN, which operate on the same frequency band as this product, there is a possibility that
interference may occur. If interference occurs, stop the operation of the other devices or relocate this product away from
other wireless devices before attempting to use it.
11. Error Messages
Measured temperature too high
Displays «H» when measured temperature is higher than 99.9°C /211.9°F in object mode.
Measured temperature too low
or Displays «L» when measured temperature is lower than 0.1 °C / 32.2 °F in object mode.
Ambient temperature too high
and object model.
Ambient temperature too low
5.0 °C / 41.0 °F in object mode.
Error function display
«Err»: The system has a malfunction.
«Errd»: Device is over 5cm from forehead /object.
«Er2»: Keep the measuring distance of 1-5 cm. Do not touch the bottom side (sensing area) of the measuring sensor.
23 : Check if the batteries have been inserted correctly. Also check polarity (<+> and <->) of the
Blank display
batteries.
18 : Displays «ErrH» when measured temperature is higher than 43°C/109.4°F in body mode or
19 : Displays «ErrL» when measured temperature is lower than 34.0 °C / 93.2 °F in body mode
20 : Displays «AH» when ambient temperature is higher than 40.0 °C / 104.0°F in body model
21 : Displays «AL» when ambient temperature is lower than 15.0 °C / 59.0 °F in body mode and
22 :
appears steady on the display.
14