Flat battery indicator
immediately.
12. Cleaning and Disinfecting
Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer casing and the
measuring sensor. Ensure that no liquid enters the interior of the device. Never use abrasive cleaning agents, thinners or
benzene for cleaning and never immerse the device in water or other cleaning liquids. Take care not to scratch the
surface of the sensor lens and the display.
13. Battery Replacement
This device is supplied with 2 new, long-life 1.5V, size AAA batteries. Batteries need replacing when this icon
«battery»
14 is the only symbol shown on the display.
Remove the battery cover
by the symbols in the compartment.
Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with
domestic waste.
14. Guarantee
This device is covered by 5 years guarantee from the date of purchase. The guarantee is valid only on presentation of
the guaranteed card completed by the dealer (see back) confirming date of purchase or the receipt.
The guarantee covers the device. Batteries and packaging are not included.
Opening or altering the device invalidates the guarantee.
The guarantee does not cover damage caused by improper handling, discharged batteries, accidents or
non-compliance with the operating instructions.
Please contact Microlife-service.
15. Technical Specifications
Type:
Measurement Range:
Resolution: 0.1 °C / °F
Measurement accuracy (Laboratory): Body mode: ± 0.2°C (35.0°C~42.0°C) / ± 0.4°F (95.0°F~107.6 °F)
Liquid Crystal Display, 4 digits plus special icons
Display:
24 : If only «battery» icon is shown on the display, the batteries should be replaced
7 by sliding it in the direction shown. Replace the batteries – ensure correct polarity as shown
Infrared Forehead Thermometer FR1DQ1-B
Body mode: 34.0 °C -43.0 °C / 93.2 °F -109.4 °F,
Object mode: 0.1-99.9 °C / 32.2 °F-211.8.0 °F
± 0.3°C (34.0°C~34.9 °C and 42.1°C~43.0°C) / ± 0.5°F(93.2°F~94.8 °F and 107.8°F~109.4°F)
Object mode: ± 1.0°C (0.1°C~99.9°C) / ± 2 °F (32.2°F~211.8 °F)