Replacing The Overpressure Safety Relief Device - vacuubrand MZ 2C NT+2AK Instructions For Use Manual

Chemistry diaphragm pumps
Table of Contents

Advertisement

Replacing the overpressure safety relief device

page 67 of 80
Overpressure
safety relief device .....................................20638821
(at exhaust waste vapor condenser)
➨ Remove catchpot (9) at outlet (see "Use
and operation", pg. 40).
➨ Loosen the union nut at the inlet of the ex-
haust vapor condenser.
➨ Remove the 4 screws affixing the counter
holder of the exhaust vapor condenser with
a Torx driver T10. Remove the exhaust va-
por condenser. In doing so pull the PTFE
hose out of the inlet of the exhaust vapor
condenser.
➨ If necessary, pull off the old pressure relief
valve and install the new one. Check for
correct position of the PTFE-foil under the
pressure relief valve.
➨ Feed the PTFE hose into the inlet of the
exhaust vapor condenser. Mount the con-
denser with the counter holder (Torx screws
T10) to the pump. Fasten union nut.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents