Technical Data; Evaluation Unit F 5673 - BARTEC BENKE HYGROPHIL HCDT Series Operating Manual

Measuring system for determining the dew point of hydrocarbons in natural gas
Table of Contents

Advertisement

1.4
1.4.1
Evaluation unit
Specific data
Display ranges
(not the measuring range)
Display
Keyboard
Electrical data
Auxiliary energy
Auxiliary variables
Measurement channels
Measuring rate
Inputs
Analog outputs
(at Analog-IO Ex Typ 5673-114)
3 analog outputs at the COM-
plug in unit
Control outputs
Interfaces
HY GROPHIL® F 5673 Operating manual, Sof tware v ersion 1.8.x, 387897MDHEN (2013/4/26)

Technical data

Evaluation unit F 5673

Measurement variable
Vol%
[%]
PPMV
[ppm]
VP
[hPa]
3
MC
[mg/m
],
[lb/MMscF]
DT/FP
[° C,° K,° F]
HCDT
[° C]
[bar, PSI,...]
SP
TT
[° C,° K,° F]
Graphic display with 320 x 240 dots
Red backlit touch-screen
DC 10 - 36 V max. 60 W (Si 6.3 A T)
AC 100 - 240 V max. 110 VA (Si 3.15 A MT)
The temperat ure (TT) at the measuring point is measured via a sensor int e-
grated in the humidity sensor or entered by hand. The pressure (SP ) is
measured via a 4...20 mA signal from an external pressure transmitter or en-
tered by hand. CO
-content (0...100%) is measured via a 4...20 mA-signal or
2
entered by hand.
1 x DT, 1 x HCDT
Max. 3 measurements per minute (DT)
6 measurements per hour at 5 ° C HCDT
(first measuring value after 30 min)
- Light waveguide port for optical humidity sensor
- 9-pole D-SUB plug for for characteristic curve of sensor
- Clamp-type terminal for PT 100 measurement sensor in sensor head
- Clamp-type terminal for signal from a pressure transmitter for absolute
pressure (4...20 mA)
- Clamp-type terminal for HCDT Sensor
- Clamp-type terminal for CO
all inputs Ex ia, galvanically isolated
Clamp-type terminal 0/4...20 mA, EEx ia, galvanically isolated
Source or sink
intrinsically safe
Resolution 0.0003 mA
Max. load: 500 Ω
Accuracy: 0.03 mA
Temperature drift: <0.001 mA/ ° C
Clamp-type terminal 0/4...20 mA, galvanically isolated
Source
not intrinsically safe
Resolution 0.0003 mA
Load <800 Ω
Accuracy ± 0.15 % (0.03 mA)
Temperature drift < 0.001 mA/° C
8 relay changeover contact, 30 V / 1 A
Connection via plug-in terminals
2 switching outputs "Limit" and "Error"
Ethernet, RS 232, RS 485 Modbus, Profibus, USB
System description
技术参数
F5673 电子单元
Lower range limit
0
1
0
0.5
-100
-30
0
-50
content (4...20 mA)
2
系统介绍
1-13
Upper range limit
100
25000
250
30000
+100
+30
250
+100